2.4 Signal Outputs
Local Display:
Digital display calibrated for oxygen and carbon dioxide.
The range is stated on the serial number and can be accessed via the joystick on the
front panel. In the measurement mode pushing the joystick down will scroll the gas
and range on the display. Push the joystick down again to stop the scrolling and
display the gas again.
Standard Analog Output: DC 4-20 mA
Optional Relay Output: Dual level user selectable alarm relays and one fault relay.
Rated, 2amps @ 24VAC or 24VDC
2.5 Electrical Requirements
Power:
24 VDC external power. A regulated 24VDC power supply is required.
Consumption:
Approximately 250mA
2.6 Physical Characteristics
Dimensions:
6.5 (W) x 3.15 (H) x 3.00 (D) inches; 165 x 80 x 76 mm (Max with feet)
Weight:
1.1 pounds (0.5 kg)
Enclosure Type: General purpose; not intended for explosive atmospheres.