Annexes
Annex 1: Selected error messages of the device
Error
Description and debugging
Error 1
A/D converter is at low limit, 0x0000.
Contact support.
Error 2
Temperature probe is not connected, or cable is damaged.
Error 3
Internal converter communication error.
Contact support.
Error 5
Few valid measured values for calculating the average value
for the recording interval (more than 1/8 of invalid values were
measured from the given recording interval).
Error 9
Measured value is out of range or probe is shorted.
Error 10
The device is switched off, measured values are not available.
Error 11
No valid data of the counter, batteries missing in the device.
Reset the counter content by means of the SW, repeat
the measurement.
Error 13
Error 14
Error 15
CO
2
concentration sensor is not available, or it does not work correctly.
Have the device repaired. This applies to the CO
2
concentration
sensor devices only.
Error 16
The measured CO
2
value is not available due to low battery voltage.
Recharge the battery.
Error 17
The digital T / RH probe does not communicate with the instrument.
Check its connection.
Error 18
The digital T/RH probe memory does not communicate with the
instrument.
Contact support
.
Error 19
Measured value is not available.
In device U3121M (U3121G) check, and, if needed, replace the
Digi/E probe connected. In the other models a device failure may be
suspected.
Summary of Contents for U0110G
Page 5: ...IE LGR UxxxxM 23 5 Dataloggers UxxxxM and UxxxxG U0141M U0141G models...
Page 61: ...IE LGR UxxxxM 23 61 Dimensions U0110M U0110G U8410M U8410G U0121M U0121G...
Page 62: ...62 IE LGR UxxxxM 23 U0141M U0141G U3120M U3120G U4440M U4440G...
Page 63: ...IE LGR UxxxxM 23 63 U3121M U3121G U3631M U3631G...
Page 64: ...64 IE LGR UxxxxM 23 U0141TM U0141TG U0843M U0843G U5841M U5841G U6841M U6841G U7841M U7841G...