Comet T4311 Instruction Manual Download Page 9

IE-SNC-T43(4)11-02 

 

  9 

is sent as eight bit data word in format: 1 start bit, 8 bit data word (LSB first), 2 stop bits, no parity. 
Transducer supports communication speed from 110Bd to 115200Bd. 
Sent request and response have syntax: ADDRESS OF DEVICE – FUNCTION – Modbus CRC 

Supported functions 

03 (0x03)

: Reading of 16-bit registers (Read Holding Registers) 

04 (0x04)

: Reading of 16-bit input gates (Read Input Registers) 

16 (0x10)

: Setting of more 16-bit registers (Write Multiple Registers)*

2

 

Jumper and button 

Jumper and button are located next to connection terminals. If Modbus communication protocol is 
used, the function of jumper and button is as follows: 

 

Jumper opened – transducer memory is write protect, from transducer is only enabled to 
read measured value, writing to memory is disabled (no change of transducer address and 
communication speed is enabled) 

 

Jumper closed – writing to transducer memory is enabled by means of function 

16 (0x10): 

Setting of several 16-bit registers (Write Multiple Registers)

 

 

Jumper closed and pressed button for longer than three seconds – causes restoring of 
manufacturer setting of communication protocol, i.e. sets transducer address to 01h and 
communication speed to 9600Bd 

 

Jumper opened and pressed button – transducer goes to Info mode, see chapter „

Info mode

“ 

Description of supported function 

03 (0x03): Reading of 16-bit registers (Read Holding Registers) 

Function serves for reading of values from transducer. Addresses of available registers are listed 
in chapter „

Modbus registers of the

 transducer at page 

14

.“ 

 
Request: 
---------------------------------------------------------------------------------------------- 
FUNCTION 

Function 

code 

  0x03 

---------------------------------------------------------------------------------------------- 
DATA  

Initial address Hi 

 

0x?? 

  Initial 

address 

Lo 

 0x?? 

---------------------------------------------------------------------------- 

 

 

Number of registers Hi 

0x?? 

 

 

Number of registers Lo 

0x?? 

---------------------------------------------------------------------------------------------- 
 
Response: 
---------------------------------------------------------------------------------------------- 
FUNCTION 

Function 

code 

  0x03 

---------------------------------------------------------------------------------------------- 
DATA  

Number of Bytes 

 

0x?? 

---------------------------------------------------------------------------- 

 

 

States of register Hi   

0x?? 

 

 

States of register Lo   

0x?? 

  ... 
 

 

States of register Hi   

0x?? 

                                                 

2

 See detailed description of this function 

Summary of Contents for T4311

Page 1: ...TRANSDUCER T4311 T4411 Programmable temperature transducer for RTD Pt1000 sensor with serial output RS232 and RS485 Instruction manual...

Page 2: ...le Registers 10 Exception Responses 13 Exception codes 13 Modbus CRC 13 A procedure for generating a CRC is 13 Modbus registers of the transducer 14 Example of communication 14 Reading of temperature...

Page 3: ...ion of the transducer Transducer is designed for wall mounting There are two mounting holes at the sides of the case Interconnection terminals are accessible after unscrewing four screws and removing...

Page 4: ...C T43 4 11 02 The cable should not be led in parallel along power cabling Safety distance is up to 0 5 m otherwise undesirable induction of interference signals can appear Dimensions T4311 RS232 T4411...

Page 5: ...IE SNC T43 4 11 02 5 Typical application wiring connection of terminals T4311 power from communication port T4311 power from external voltage source T4411 RS485...

Page 6: ...s Operating temperature range 30 to 70 C with switched OFF display up to 80 C Operating relative humidity range 0 to 100 RH Outer influence in accordance with EN 33 2000 3 normal environment with thos...

Page 7: ...COM1 and press OK In following window COM1 Features enter to field Bits per second 9600 Data bits 8 Parity None Number of stop bits 1 Flow control None Continue by pressing OK Select item File Featur...

Page 8: ...communication to default value from manufacturer Modbus address 01h 9600 Bd 9 End of setting All changes in setting are continuously automatically stored If needed close transducer again with cover D...

Page 9: ...ion speed is enabled Jumper closed writing to transducer memory is enabled by means of function 16 0x10 Setting of several 16 bit registers Write Multiple Registers Jumper closed and pressed button fo...

Page 10: ...tting of transducer address and communication speed is possible to perform by writing to transducer registers Attention During write to transducer registers it is not enabled to write any number of re...

Page 11: ...is sum it enables to check correct reading of the area 3 Modify content corresponding to addresses 0x2001 and 0x2002 in master device as required Setting of other values should not be changed 4 Calcul...

Page 12: ...12 ED BF 77 D5 4F 10 77 D8 FF FF FF FF 40 DE 77 D3 2E F7 78 0C 06 5C 00 01 00 00 00 00 F3 DC 00 12 42 9F 52 3A 00 9F new address of the transducer original address was 00 01 00 24 new code of corresp...

Page 13: ...code ad2 Data field Into exception response transducer returns exception code in data field Event causing exception is determined this way Exception codes 0x01 Invalid function Function code in query...

Page 14: ...Write Multiple Registers In case there is a need for reading of measured values from transducer with higher resolution than one decimal measured values in transducer are stored also in Float format wh...

Page 15: ...ransducer always communicates with following parameters without regardless of setting stored in transducer communication speed 9600 Bd no check sum transducer address 00h If jumper during switching ON...

Page 16: ...mediately after jumper is opened If address is changing o and jumper is closed transducer responds with address 00h again and newly set address will be activated after jumper is opened o and jumper is...

Page 17: ...etting of communication speed and CRC must not change setting of communication speed and CRC is not possible to change without closed jumper Command 23242B0600 cr Response 24 cr Example 2 Reading of t...

Page 18: ...rument to fix Error 1 Measured value is over upper limit of allowed full scale range There is a reading Err1 on LCD display Value read from the transmitter is 999 9 This state appears in case of Measu...

Page 19: ...ied temperature probe model Resolution 0 1 C Accuracy without probe 0 2 C from 100 to 100 C 0 2 from reading over above range Operating temperature range transducer 30 to 80 C external probe depends o...

Page 20: ...ory for connection of transmitter transmitters with RS485 interface to the PC via serial port RS232 Connect connector marked RS232 directly to the PC connect power to connector marked RS485 Power volt...

Reviews: