IE-SNC-T43(4)11-02
7
Setting of transducer, selection of communication protocol
Each new setting of transducer parameters is stored in its internal memory
1
. If modification of
transducer setting is needed (communication protocol, its parameters, setting of LCD display)
proceed in following way:
•
Connect transducer with PC, for T4411 (RS485) use converter RS485/RS232.
•
Run program „HyperTerminal“ on your computer – it is a part of Windows operating system
(Start
→
Program Files
→
Accessory
→
Communication
→
HyperTerminal) and set
transmission parameters. I.e. enter name of new connection (e.g. Sensor), press button OK,
window „Connect“ appears. In field „Connect by means of“ select COM port, which transducer
is connected to (e.g. COM1) and press OK. In following window „COM1 – Features“ enter to
field:
„
Bits per second
“
9600
“
Data bits
“
8
“
Parity
“
None
“
Number of stop-bits
“
1
“
Flow control
“
None
Continue by pressing OK.
•
Select item „File“, „Features“ in menu. Select bookmark „Setup“, in the window, click on
„ASCII Setup... “ and modify all items in
accordance with following figure. Then press OK
•
Unscrew 4 screws of transducer cover and
remove it. Plug in jumper.
•
Press button inside of the transducer and only
now you can connect power voltage. If transducer
is powered directly from PC communication port
(for interface RS232), it is necessary first
disconnect communication cable, then press
button, connect communication cable (and also
power) and only now release button.
•
LCD display of the transducer displays reading
„PC“.
1
During writing to transducer memory no power failure should appear! Such failure can cause
damage of other settings stored in transducer!