Comet T4311 Instruction Manual Download Page 3

IE-SNC-T43(4)11-02 

 

  3 

 

Instruction Manual for use of transducer: 

T4311 (RS232) 
T4411 (RS485) 

 

Transducer is designed for temperature measurement by means of the external temperature 

probe with RTD Pt1000 sensor. It is built in a plastic case with the IP65 protection. Read this 
manual before first transducer connection. Transducer T4311 communicates via link RS232, 
transducer T4411 via link RS485. Supported communication protocols are Modbus RTU and 
protocol compatible with standard Advantech-ADAM. Measured value is displayed on dual line 
LCD display. Display can be also switched OFF. Output link RS485 of transducer T4411 is 

galvanic isolated

. Output link RS232 of transducer T4311 is 

NOT

 galvanic isolated. Setting of all 

transducer parameters is enabled in accordance with procedure described in chapter „

Setting of 

transducer, selection of communication protocol

“ or by means of commands of selected 

communication protocol. For detailed description of features of protocols see chapter „

Description 

of communication protocols

“. 

Setting of transducer from the manufacturer 

If special setting was not required in the order, transducer is set from the manufacturer to the 
following parameters: 

 

communication protocol: 

Modbus RTU

 

address of transducer: 

01H

 

communication speed: 

9600Bd, no parity, 2 stop bits

 

display:  

switched ON

 

 

Installation of the transducer 

Transducer is designed for wall mounting. There are two mounting holes at the sides of the 

case. Interconnection terminals are accessible after unscrewing four screws and removing the lid. 
Lace the cable through a gland at the case wall. Connect the cable to terminals with respecting the 
signal polarity (see figure). Terminals are self-clamping and can be opened by a suitable 
screwdriver. For the opening, insert the screwdriver to upper terminal hole and lever by him. Do not 
remember to tighten glands and case lid with inserted packing after cables connecting. It is 
necessary for warranting of protection IP65.  

Working position is negligible. 
 

Transducer T4311 is supplied with connection cable equipped with connector for connection to 

RS232 interface. For T4411 transducer (RS485) it is recommended to use shielded twisted copper 
cable (e.g. SYKFY ) with outside diameter of the cable must be from 3 to 6.5 mm, maximal length 
1200m. The cable must be located at indoor rooms.  

Nominal cable impedance should be 100 

, loop resistance max. 240 

, cable capacity max. 

65 pF/m. Suitable cable is e.g. SYKFY 2x2x0,5 mm

2

, where one wire pair serves for transducer 

powering and second pair for communication link. 

The cable should be led in one line, i.e. NOT to „tree“ or „star“. At the end termination 

resistor should be located. For short distance other topology is allowed. Terminate the network by a 
termination resistor. The value of the resistor is recommended about 120 

. For short distance 

termination resistor can be left out. 

Summary of Contents for T4311

Page 1: ...TRANSDUCER T4311 T4411 Programmable temperature transducer for RTD Pt1000 sensor with serial output RS232 and RS485 Instruction manual...

Page 2: ...le Registers 10 Exception Responses 13 Exception codes 13 Modbus CRC 13 A procedure for generating a CRC is 13 Modbus registers of the transducer 14 Example of communication 14 Reading of temperature...

Page 3: ...ion of the transducer Transducer is designed for wall mounting There are two mounting holes at the sides of the case Interconnection terminals are accessible after unscrewing four screws and removing...

Page 4: ...C T43 4 11 02 The cable should not be led in parallel along power cabling Safety distance is up to 0 5 m otherwise undesirable induction of interference signals can appear Dimensions T4311 RS232 T4411...

Page 5: ...IE SNC T43 4 11 02 5 Typical application wiring connection of terminals T4311 power from communication port T4311 power from external voltage source T4411 RS485...

Page 6: ...s Operating temperature range 30 to 70 C with switched OFF display up to 80 C Operating relative humidity range 0 to 100 RH Outer influence in accordance with EN 33 2000 3 normal environment with thos...

Page 7: ...COM1 and press OK In following window COM1 Features enter to field Bits per second 9600 Data bits 8 Parity None Number of stop bits 1 Flow control None Continue by pressing OK Select item File Featur...

Page 8: ...communication to default value from manufacturer Modbus address 01h 9600 Bd 9 End of setting All changes in setting are continuously automatically stored If needed close transducer again with cover D...

Page 9: ...ion speed is enabled Jumper closed writing to transducer memory is enabled by means of function 16 0x10 Setting of several 16 bit registers Write Multiple Registers Jumper closed and pressed button fo...

Page 10: ...tting of transducer address and communication speed is possible to perform by writing to transducer registers Attention During write to transducer registers it is not enabled to write any number of re...

Page 11: ...is sum it enables to check correct reading of the area 3 Modify content corresponding to addresses 0x2001 and 0x2002 in master device as required Setting of other values should not be changed 4 Calcul...

Page 12: ...12 ED BF 77 D5 4F 10 77 D8 FF FF FF FF 40 DE 77 D3 2E F7 78 0C 06 5C 00 01 00 00 00 00 F3 DC 00 12 42 9F 52 3A 00 9F new address of the transducer original address was 00 01 00 24 new code of corresp...

Page 13: ...code ad2 Data field Into exception response transducer returns exception code in data field Event causing exception is determined this way Exception codes 0x01 Invalid function Function code in query...

Page 14: ...Write Multiple Registers In case there is a need for reading of measured values from transducer with higher resolution than one decimal measured values in transducer are stored also in Float format wh...

Page 15: ...ransducer always communicates with following parameters without regardless of setting stored in transducer communication speed 9600 Bd no check sum transducer address 00h If jumper during switching ON...

Page 16: ...mediately after jumper is opened If address is changing o and jumper is closed transducer responds with address 00h again and newly set address will be activated after jumper is opened o and jumper is...

Page 17: ...etting of communication speed and CRC must not change setting of communication speed and CRC is not possible to change without closed jumper Command 23242B0600 cr Response 24 cr Example 2 Reading of t...

Page 18: ...rument to fix Error 1 Measured value is over upper limit of allowed full scale range There is a reading Err1 on LCD display Value read from the transmitter is 999 9 This state appears in case of Measu...

Page 19: ...ied temperature probe model Resolution 0 1 C Accuracy without probe 0 2 C from 100 to 100 C 0 2 from reading over above range Operating temperature range transducer 30 to 80 C external probe depends o...

Page 20: ...ory for connection of transmitter transmitters with RS485 interface to the PC via serial port RS232 Connect connector marked RS232 directly to the PC connect power to connector marked RS485 Power volt...