Operating conditions:
Operating temperature range: -30 to +80 °C, over +70 °C switch LCD display OFF
external probe - depends on probe model.
Operating humidity range: 0 to 100 %RH
Outer characteristics in accordance with Czech National Standard 33-2000-3:
normal environment with the specifications: AE1, AN1, BE1
Working position: arbitrary
Electromagnetic compatibility: complies EN 61326-1
Not allowed manipulations: It is not allowed to operate the device under other than
specified conditions in technical parameters. Devices are not designed for locations
with chemically aggressive environment.
Storing conditions: temperature -30 to +80 °C
humidity 0 to 100 %RH without condensation
Dimensions: see dimensional drawings
Weight: approximately 145 g
Material of the case: ASA
Procedure of modification of transducer adjustment:
•
Device adjustment is performed by means of the optional SP003 communication cable,
connected to USB port of the PC.
•
It is necessary to have installed configuration program Tsensor on the PC. It is free to
download at
. During installation please take care about
installation of driver for USB communication cable.
•
Connect SP003 communication cable to the PC. Installed USB driver detect connected
cable and create virtual COM port inside the PC.
•
Unscrew four screws of the device lid a remove the lid. If device is already installed to
measuring system, disconnect leads from terminals.
•
Connect SP003 communication cable to the device. Display must light up, or at least
must light up all symbols for one second (if LCD was switched OFF by program
before).
•
Run installed Tsensor program and select corresponding communication COM port (as
described above).
•
When new setting is saved and finished, disconnect the cable from the device, connect
leads into its terminals and place the lid back to the device.
Error States of the device
Device continuously checks its state during operation. In case error is found LCD displays
corresponding error code:
Error 0
First line displays „Err0“.
Check sum error of stored setting inside device’s memory. Output value is < 3.8 mA. This error
appears if incorrect writing procedure to device’s memory occurred or if damage of calibration data
appeared. At this state device does not measure and calculate values. It is the serious error, contact
distributor of the instrument to fix.