Comet T4111 Instruction Manual Download Page 5

IE-SNC-T4111-06 

 

  5 

 
Operating conditions: 

Operating temperature range: -30 to +80 °C, over +70 °C switch LCD display OFF 

external probe - depends on probe model. 

Operating humidity range: 0 to 100 %RH 
Outer characteristics in accordance with Czech National Standard 33-2000-3:  

normal environment with the specifications: AE1, AN1, BE1 

Working position: arbitrary 
Electromagnetic compatibility: complies EN 61326-1  
Not  allowed  manipulations:  It  is  not  allowed  to  operate  the  device  under  other  than 

specified conditions  in  technical  parameters. Devices  are not  designed for locations 
with chemically aggressive environment.  

Storing conditions: temperature -30 to +80 °C 

 humidity 0 to 100 %RH without condensation 

Dimensions: see dimensional drawings

 

 

Weight: approximately 145 g 
Material of the case: ASA 
 
 

Procedure of modification of transducer adjustment: 

 

Device adjustment is performed by means of the optional SP003 communication cable, 
connected to USB port of the PC. 

 

It is necessary to have installed configuration program Tsensor on the PC. It is free to 
download  at 

www.cometsystem.com

.  During  installation  please  take  care  about 

installation of driver for USB communication cable. 

 

Connect SP003 communication cable to the PC. Installed USB driver detect connected 
cable and create virtual COM port inside the PC. 

 

Unscrew four screws of the device lid a remove the lid. If device is already installed to 
measuring system, disconnect leads from terminals. 

 

Connect  SP003  communication  cable  to  the  device.  Display  must  light  up,  or  at  least 
must  light  up  all  symbols  for  one  second  (if  LCD  was  switched  OFF  by  program 
before).  

 

Run installed Tsensor program and select corresponding communication COM port (as 
described above). 

 

When new setting is saved and finished, disconnect the cable from the device, connect 
leads into its terminals and place the lid back to the device. 

 

Error States of the device 

Device  continuously  checks  its  state  during  operation.  In  case  error  is  found  LCD  displays 

corresponding error code: 

Error 0 

First line displays „Err0“. 
Check sum error of stored setting inside device’s memory. Output value  is < 3.8 mA. This error 
appears if incorrect writing procedure to device’s memory occurred or if damage of calibration data 
appeared. At this state device does not measure and calculate values. It is the serious error, contact 
distributor of the instrument to fix. 