Page 24 of 39
INSTRUCTION MANUAL
SECT IO N 5 -
CALIBRATION OF ELECTRICAL PERFORMANCE PARAMETERS
1912 1 E l T o ro R d
●
Si l vera do, Ca lifo rni a 9 26 76
●
(9 49) 459 -96 00
●
Rev050321
Using the S-Parameter data and the ABCD Transformation Equations
given in section 5.1.2, calculate the Decoupling Factor of the EM
clamp at each test frequency using the equation given below.
No limits are defined for decoupling factor. Typical CLEM-6146 EM
Clamp data are shown in Figure 6 of the present document.
Typical decoupling factor performance curves for three (3) different
clamps are provided in Figure A.8 in Edition 4.0 of IEC 61000-4-6.
Using the S-Parameter data and the ABCD transformation
equations given in Section 5.1.2, calculate the input impedance of
the EM Clamp at each test frequency using the equation given
below.
No limits are defined for the input impedance. Typical CLEM-6146
EM Clamp input impedance data are shown in Figure 7 of the
present document.
Typical decoupling factor performance curves for three (3) different
clamps are provided in Figure A.7 in Edition 4.0 of IEC 61000-4-6.
Distributed by: Reliant EMC LLC, +1 408 916
‐
5750, [email protected], www.reliantemc.com