11.6 Standard Accessories
(1)
Disposable Adult Cuff
(1)
Adult Cuff (12 cm in width)
(1)
3.5 m pressure hose
(2)
Paper Roll (for units with printers)
(1)
Paper Shaft (for units with printers)
(1)
AC-power cord
(2)
Boxes Patient Pack Genius Probe Covers
(1)
Rolling Stand
(1)
Dust Cover
(1)
Operations Manual
11.7 Communication Specifications
a.
Asynchronous
b.
4800 bps
c.
1 start bit
d.
8 data bits
e.
Even Parity
f.
1 Stop Bit
g.
Ready/Busy Control by RTS/CTS Signal
Data can be received by Press-Mate when RTS signal level is HIGH. Data can be transmitted
by Press-Mate when CTS signal level is HIGH.
h.
RS-232C ouputs data when Temp-Mate is connected.
i.
When a measurement is completed, data is output according to the format in the following
diagrams:
Ø
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Ø
YEAR
MONTH
DAY
SP
HOUR
:
MINUTE
SP
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
SYS
SP
MEAN
SP
DIA
SP
PR
SP SP
ALARM MARK
33
74
75
76
ALL SP CR LF
"/"
= ASCII code 2 FH
":"
= ASCII code 3 AH
"SP"
= ASCII code 2 ØH
"CR"
= ASCII code Ø D
H
"LF"
= ASCII code Ø A
H