6 .3 Calibration Table Commands and Queries
CTSTORE arg1 arg2
Store one entry into calibration table
CTSTORE on | off
Use calibration table command
CTSTORE save
Save calibration table to NVRAM
CTSTORE load
Load calibration table to memory
CTSTORE info
Report calibration table information
CTSTORE reset
Reset and clear calibration table
CTSTOREM
Store multiple entries into calibration table
CTSTORE?
Use calibration table query
CTSTORE? arg1
Query individual entries in calibration table
CTSTORE
Calibration table parameter set (COMMAND)
Description
Operates on the in-memory calibration table. The in-memory calibration table (when enabled) will apply a
calibration position offset to ensure positional accuracy of the stepper motor.
Syntax
CTSTORE [ save | load | reset | on | off | info [arg1]]
save
Stores in-memory calibration table into non-volatile random-access memory
(NVRAM).
load
Load into memory the calibration table from NVRAM.
reset
Clears all in-memory calibration table entries. Sets “no calibration” text entry into
the info setting.
on
use Calibration table compensation (default)
off
do not use Calibration table compensation when setting delay settings
info
set the Calibration table text entry to [arg1]. 128x characters max.
Returns
None
CTSTORE save
Calibration table save (COMMAND)
Description
Saves calibration table to non-volatile memory
Returns
None
CTSTORE load
Calibration table load (COMMAND)
Description
Load calibration table from non-volatile memory
Returns
None
22
COLBY INSTRUMENTS XT-100 VERSION 1.0 OPERATING AND PROGRAMMING MANUAL