A.1 Firmware Update Instructions
WARNINGS
■
The firmware update process will erase all job and settings files from In-Sight non-volatile flash memory.
In-Sight Explorer creates a backup of all job files and settings prior to the firmware update. These files are
restored to the wafer reader automatically after the firmware update; however, you may wish to perform a
manual backup operation on the wafer reader before starting the Firmware Update procedure.
■
Do not power cycle the wafer reader while the firmware is being upgraded.
■
If an error occurs during the firmware update process, power cycle the In-Sight wafer reader and restart the
firmware update.
■
You cannot update the firmware on a wafer reader that is currently Online. Take the wafer reader Offline
before attempting to update the firmware.
■
If a job containing a large number of functions is loaded when a firmware update is attempted, the update may
be interrupted due to lack of empty flash memory. If this occurs: (1) ensure that the wafer reader is Offline, (2)
click File > New Job and (3) try the update again.
■
If the jobs on an In-Sight wafer reader consume a large portion of available flash memory, the firmware
update may be interrupted due to lack of empty flash memory. If this occurs: (1) backup or delete the job files
from the wafer reader, and (2) try the update again.
■
If you receive an error ("Error 6029: Unable to establish native mode connection to <wafer reader>") while
attempting the firmware update process, you may be logged on to the wafer reader without the appropriate
privileges. Log off and then log on as a user with appropriate privileges and retry the firmware update
procedure.
A.1.1 Updating the Firmware
1. Launch In-Sight Explorer.
2. Verify the following:
All target wafer readers are Offline
No other users are logged on to the target wafer readers
A new job is loaded on each target wafer reader (File > New Job)
3. On the System menu, click Update Firmware.
4. Highlight the In-Sight wafer reader(s) to be updated.
5. Click
Update
Firmware.
6. Verify that each wafer reader listed in the dialog is to receive new firmware.
7. Click
OK.
8. Once all wafer readers have been updated and rebooted, click OK to close the
Firmware Update dialog.
Summary of Contents for In-Sight 1720 Series
Page 1: ...1 ...
Page 2: ......
Page 6: ...Precautions iv ...
Page 10: ...List of Figures viii ...
Page 12: ...List of Tables x ...
Page 22: ...Set Up the PC 10 ...
Page 54: ...Connecting the Breakout Module 42 ...
Page 56: ...P N 597 0047 01 www cognex com Installing the In Sight 1720 Series Wafer Reader ...