7.9.7. Parameters
Select this option by pressing the
ENTER
key. The
and
keys will now display the available options from
within this sub-mode, when the option that requires changing is displayed, press the
ENTER
key. When all
required changes have been made, select the
EXIT
option and press
ENTER
.
5 PARAMETERS
7.9.7.1. Security Number
To prevent any unauthorised tampering with the set up information, it is important that the security code is
changed from the factory setting. Each digit is selected with the
ENTER
key and changed with the
and
keys.
5 PARAMETERS
Security # 0000
It is important to make a note of this number otherwise it will not be possible to change the
instrument set-up.
7.9.7.2. Units
The system operates in metric units to provide output data in m/sec.
5 PARAMETERS
Units metric
7.9.7.3. Detector Separation (Distance)
The current value entered for separation will be displayed for 1 second, then the display will default to zero. The
separation must be re-entered for the unit to calculate flow correctly.
5 PARAMETERS
Distance 1000mm
7.9.7.4. Cross Sectional Area
The cross sectional area must be entered in m
2
to enable volumetric flow to be correctly displayed.
5 PARAMETERS
CS Area 010.0m2
7.9.7.5. Data Rate
The
DATA RATE
selected will depend on the range of flow velocities to be measured. Estimate the Time of
Flight (TOF) for the minimum and maximum expected flow velocities as follows:
Maximum TOF (ms) = separation (mm)/minimum flow velocity (m/s)
Minimum TOF (ms) = separation (mm)/maximum flow velocity (m/s)
The
DATA RATES
available are 0.5, 1.0, 2.0, 4.0 and 8.0ms.
Select the
DATA RATES
such that:
Max.
TOF/DATA RATE
255
Min.
TOF/DATA RATE
50