Issue : B
Rev. : 1
Date : 23/10/13
Doc. i/d : 0109/6
Ref. : 090028
Press the
ENTER
key when this display is shown, the display will now show one of the averages. Use the
and
keys to select the average time that requires changing, and press the ENTER key to change it. The value
can now be changed using the
and
keys and input by pressing the
ENTER
key.
5 SET AVERAGES
Set the seconds averaging stack to the required value. This is limited to within 10 to 60 seconds in 10-second
intervals.
5 SET AVERAGES
secs 60
Set the minutes averaging stack to the required value. This is limited to within 1 to 60 minutes in 1-minute
intervals.
5 SET AVERAGES
mins 60
Set the hours averaging stack to the required value. This is limited to within 1 to 24 hours in 1-hour intervals.
5 SET AVERAGES
hours 24
Set the days averaging stack to the required value. This is limited to within 1 to 30 days in 1-day intervals.
5 SET AVERAGES
days 30
7.9.3. Configure O/P1
The analogue current loop output (terminals 19 & 20), is set up from this mode. Press the
ENTER
key while this
display is shown to select it, then press the
and
keys to step through the available options. Press the
ENTER
key to enter each option and change the displayed parameter.
5 CONFIGURE O/P1
7.9.3.1. Base of Output
An origin of 0 or 4mA can be set for the current loop output. The
and
keys will toggle between these two
options. Press the
ENTER
key to enter the new value.
5 CONFIGURE O/P1
Output = 4 to 20mA