Page: 68
EnergyTech 401
CODEL
Operating and Maintenance Manual
Monitoring Solutions
Current Output Calibration
4.10
The current output should now be set up using a calibrated milliammeter set to
DC current, 20mA max. Conduct this procedure as follows :
Connect the milliammeter to the output terminals within the signal processor –
termA & -mA.
Enter the SET UP MODE – CONFIGURE OUTPUT – SET ZERO OPTION, and adjust
the level using the ARROW keys until 0mA is recorded. Record the value in
brackets on the display in Table 2 – Calibration Data.
Enter the SETSPAN option and adjust the level using the ARROW keys until 20mA
is recorded. Record the value in brackets on the display in Table 2 – Calibration
Data.
The current output is now calibrated.
This completes the commissioning of the EnergyTech 401 gas analyser.
This is set at the factory and should not be altered without due consideration.
WARNING!!!