2201R Mobile Tester SCPI Reference Guide
Chapter 6 Status commands
47090/323
Copyright © 2016 Aeroflex Limited. All rights reserved.
66
:STATus:OPERation:NTRansition
Syntax
:STATus:OPERation:NTRansition <int1>
Parameters
int1 is an integer. The minimum value for int1 is 0, the maximum value is 32767. The default value is 0.
Description
Sets the NTR mask of the General operation register group. This mask is ANDed with the up to 15 bits of the NTR
transition filter of the general operation condition register to allow a reset (negative transition) of any bit (i.e. a
transition from logic "1" to "0") to reach the general operation event register.
Notes
- The default of this mask is 0 — that means that the mask does not allow any negative transition of the lower 15
bits of the condition-type register to reach the event-type register.
- Bit 16 of the general operation condition register is not used.
Query
There is no query form of this command available.
Example
:STATus:OPERation:NTRansition 32767
This means that all of the negative transitions of the condition-type register are forwarded to the event-type register.