2201R Mobile Tester SCPI Reference Guide
Chapter 8 Configuration commands
47090/323
Copyright © 2016 Aeroflex Limited. All rights reserved.
131
:CONFigure:GPRS:BER:RSIGnalling:BITPattern
Syntax
:CONFigure:GPRS:BER:RSIGnalling:BITPattern <PredefExp>
Parameters
PredefExp is one of the following predefined expressions: PRBS9|PRBS15|PRBS23|.
Default is PRBS9.
Description
This command selects the bit sequence to be transmitted by the RF generator in reduced signalling GMSK mode.
Query
The query form of this command returns the current setting. The string delivered back contains one of short-form
versions of the predefined expressions explained above.
Example
:CONF:GPRS:BER:RSIG:BITP PRBS15
:CONF:GPRS:BER:RSIG:BITP? Value returned: "PRBS15".
:CONFigure:GPRS:BER:RSIGnalling:COUNt
Syntax
:CONFigure:GPRS:BER:RSIGnalling:COUNt <int1>
Parameters
int1 is an integer parameter.
The minimum value for int1 is 100, the maximum value is 10000. The default value for int1 is 100.
Description
Defines the number of samples (blocks) for each GPRS reduced signaling BER measurement.
Query
The query form of this command returns the current setting. The string delivered back contains one integer.
Example
:CONF:GPRS:BER:RSIG:COUN 2000
:CONF:GPRS:BER:RSIG:COUN? Value returned: 2000.