Gather the following items:
  •  Coaguchek XS Pro Meter
  •  Container of Test Strips
  •  Test Strip Code Chip
  •  Safe-T-Pro Plus Lancet
  •  Alcohol Wipe
  •  Cotton Ball or Tissues

1.

Make sure you have the code chip 
ready to insert if you are opening 
a new box of strips. The code key 
needs to be inserted into the meter 
everytime you use a new lot of strips.

2.

Clean the fingertip with an alcohol 
wipe. Or have the patient wash his or 
her hands in warm soapy water.
Dry thoroughly.  

3.

Before continuing, review these 

tips for getting a good drop of 

blood.

Increasing the blood flow in the 
finger will help you to get a good 
drop of blood so keep in mind these 
tips:

Warm the hand. Have the patient 

• 

hold it under his or her arm, and/  
or wash with warm water.
Have the patient let that arm 

• 

hang by his or her side.
Massage the finger from its base.

• 

Use these techniques until the 
fingertip has increased in colour.

 

 

4.

Take a test strip out of the container.

Close the container tightly.

You have 10 minutes to use the test 
strip once you remove it from the 
container.

5.

Slide the test strip into the test strip 
guide in the direction of the arrows 
until it stops.

The meter turns on and the main 
menu appears.

If you insert a strip from a new lot 
number for the first time, the meter 
will display the number of the test 
strip code chip required.

  

Remove 

the old code chip and insert the new 
one.

6.

Touch the 

Scan 

button and hold the 

operator ID barcode approximately 
10cm from the built-in scanner. The 
meter beeps once the barcode has 
been successfully read.

7.

Press 

Patient Test 

on the main menu 

screen.

8.

Touch the 

Scan

 button and hold the 

patient ID barcode approximately 
10cm from the built-in barcode 
scanner. The meter beeps once the 
barcode has read successfully.

An hour glass symbol appears.

 

The test strip symbol appears after a 
few seconds with a flashing drop of 
blood. You have 180 seconds to apply 
the sample.

 

9.

Twist the protective cap off the 
lancet.

Massage the finger until you see 
increased colour in the fingertip.

Keeping the hand down, press the tip 
of the lancet firmly against the side of 
the fingertip. Press the purple trigger 
button.

Gently squeeze from the base of the 
finger to develop a drop of blood.

10.

Within 15 seconds of lancing the fingertip, 
apply the blood to the target area on the strip.
Find the clear target area on the test strip. You 
can dose from the side or the top.

Hold the blood drop to the test strip until you 
hear a beep. The flashing blood drop symbol 
will disappear and an hourglass symbol appears 
and the test begins.

Do not add more blood to the test strip. 

Do not touch the test strip while the test 

is in progress.

The meter performs an automatic quality 
control check on the strip before it displays 
the result.

 

QC  

  appears in the display briefly following a 

successful QC check. The result appears in less 
than a minute.

11.

Record your result as instructed by 
your facility.

Note: If during testing the meter 
displays an error message, refer to 
the Error Messages section of the 
user manual for an explanation  and 
steps on how to proceed.

12.

 Quick Reference Guide

CoaguChek  

XS Pro

®

 

P O I N T   O F   C A R E

T E S T I N G

Operator’s Manual

CoaguChek  

XS Pr

o

®

 

P O

I N

T  

O F

  C

A R

E

T E

S T

I N

G

Operator’s 

Manual