CME Medical Bodyguard 545 Quick User Manual Download Page 1

ColourVision Bodyguard

Pain Management Pump

Quick user guide

CME Medical Device Training Is RCN-Accredited 

Keypad Lock

With the keypad lock activated, the user can STOP and START an infusion, and with the infusion running use 

the INFO key to review infusion status.

Users cannot (for example) power off, rate change, change bolus or lockout time, deliver a clinician activated 

bolus or go to the main menu with the keypad lock activated.

Accessing infusion summary

During an infusion, repeatedly press the INFO to access current infusion information:

volume infused/to be infused (VI/VTBI), battery level, Pt bolus attempts/given, Clinican Bolus, protocol review 

screen and date & time

Note

o  Refer to Operating Manual for full operating instructions

o  Users must have undertaken training before operating this device

o  Screen information/sequences may vary with different software versions and local pump configuration
Always follow screen prompts and before pressing keys to proceed, ensure selections made  

correspond with what is required

CME

 

Medical UK Ltd

Kincraig Business Park,

Kincraig Road,

Blackpool FY2 0PJ

Tel: 01253 894646

Fax: 01253 896648

[email protected]

www.cmemedical.co.uk

alerts, alarms and troubleshooting

When an 

alErT

 is activated:

The infusion continues, LED indicator light remains green, intermittent beeps are heard approx. every 

3/4 mins and a screen message alternates with the infusion running screen.
When an 

alarM

 is activated:

The infusion stops, the LED indicator light turns red, a continuous audible alarm activates and a screen 

message displays indicating the cause.
Press 

STOP/NO

 to mute alarms and follow screen prompts.

Screen Description

Low Battery

Alert: Approximately 30 minutes of battery life remaining.

Connect to mains power.

Alert: Bag nearly empty, approximately 5mL volume remaining.

Prepare new bag, if applicable.

Pump has been left in the STOP state for more than 2 minutes with no key 

presses. Press START/OK to resume, press STOP/NO to continue pause or 

press and hold down STOP/NO for menu.
Battery power is depleted.

Connect to mains power immediately. Place pump in charger or use the 

battery cable to connect to the DC socket.

Current infusion programme has completed.

Either discontinue use or change bag to start a new protocol.

Air in set or the set is occluded between the infusion bag and pump.

If air is visible, disconnect set from patient and prime the set and resume 

protocol. Check for occlusion source (kinked or trapped set).
The set or access device is occluded (below pump).

Straighten the set and/or remove or open any clamp or clip. Change/flush 

the access device.
The set is loaded incorrectly or user has loaded a non-proprietary set .Load 

administration set with key in keyway. Check set is the correct dedicated 

BodyGuard set.

The pump door has opened during infusion.

Check set is loaded correctly, close door and re-start infusion.

A technical/Internal malfunction has occurred

Follow on screen prompts. User may be prompted to power off and on to 

rectify the problem.

If the problem cannot be rectified, power off and remove from patient use. 

Follow local policy and contact service dept.

Near End

Pump Stopped

End Battery

Press Stop

End of Infusion

Press STOP/NO

for Menu

Air/Up Occlusion

Check for air/

occlusion

Down Occlusion

Check set/access

Missing Set Key

Load Set Key &

Close Door

Door Open

Close Door

System Error

Press Info

Error (number)

Send for Service

CV

54