CME Medical Body Guard 323 Quick User Manual Download Page 1

323 Ambulatory Infusion Pump 

TPN Mode 

Quick user guide

Pump Alerts, Alarms and Trouble Shooting 

When an Alert Activates: 

Infusion continues, 3 beeps are heard. A message on the display 

screen alternates with the infusion running screen until the end  

of the battery or infusion. 

Low/end battery

 – the battery has approximately 30 minutes power 

remaining. Connect pump to charger or change battery pack. 

Program nearly complete

 – the programme/infusion has  

approximately 30 minutes until complete. Prepare to change  

bag and re-programme or close down the device. 

When an Alarm Activates: 

Infusion stops. LED indicator light turns red and the alarm sounds 

continuously until either the pump is paused or the problem is 

rectified. 

Always

 note the condition before pressing the ‘STOP/NO’ 

key to silence the alarm. 

Air/upstream occlusion

 – activates when either the upstream 

segment of the tubing is trapped/clamped or when more than 

the pre-set allowed volume of air has passed through the pump. 

Silence the alarm, check the upstream tubing is not occluded. 

Down occlusion

 – indicates that an occlusion has blocked the line 

or access device causing pressure build up in the set. The clamp 

may be closed or set trapped/kinked, which you should resolve. If 

successful the pump automatically resumes infusion. If you cannot 

detect and resolve the cause the access device may be blocked. 

Silence the alarm and contact your Clinician immediately. 

Refer over the page for manual and pump priming instructions. If re-

priming is indicated, ensure the line is disconnected from the patient. 

Door open

 – pump door has been opened during infusing. Simply 

close the door, silence the alarm and re-start. 

Pump unattended

 – pump has been left ‘stopped’ or unattended 

(may have programmed but forgotten to press Start). Alarm 

sounds after 2 minutes to indicate that the pump is not infusing. 

Restart pump

 – a system malfunction has occurred. Press the ‘INFO’  

key and record the error code. Some errors can be resolved by turning  

the pump off and on again. If when the pump powers on the message  

is displayed again, discontinue using the equipment, contact your 

Clinician/Provider immediately and report the error code. 

End program

 – alerts the user that the programme/infusion is  

now complete. 

Missing key

 – the administration set is loaded incorrectly.  

Check that the key is positioned correctly inside the pump  

before continuing. 

Wrong TPN Data

 – the programme entered by the user cannot be 

performed as it conflicts with either the operating parameters of the  

pump or the clinician/provider configurable settings (which require 

coded user access). Consult your clinician/provider for advice. 

Clinical contact: 
Phone number: 
E-mail contact: 

k  

The information contained in this guide is a summary only. 

Refer to Operating Manual for full operating instructions. 

k  

Screen information and sequences displayed are based on 

pump default settings and numerical values shown are 

examples only. You must refer to local policy/procedure  

for specific guidance on pump settings/set-up, protocols  

and use of accessories (e.g. cannula and administration lines). 

k  

When using the pump always follow screen prompts and 

before pressing keys to proceed ensure selections made 

correspond with what is required. 

k  

The information contained in this guide is applicable to 

software versions McTPN04, McTPN04C, McTPN05 and 

McTPN06. 

    

NOTE: With previous software versions some text wording/

screen information may vary slightly

.

CME

 

McKinley UK Ltd

 

 

Kincraig Business Park,  

Kincraig Road,  

Blackpool  FY2 0PJ 

Tel: 01253 894646  

Fax: 01253 896648  

email: [email protected]  

www.cme-mckinley.co.uk

CME McKinle

y/

323

/T

PN quick guide

/v2

/Dec 2010

CME McKinley Device Training is RCN Accredited 

The Multi-Function Keypad 

Keys ‘0’ to ‘9’ are used to enter numeric values during programming. 
Keys ‘2’ & ‘0’ are also used to scroll up and down respectively in menus. 

Powers the pump on and off. Press and hold until 

confirmation beep sounds to power off.

Starts/restarts the infusion. 

Confirms programming or selection during set up.

Stops the infusion when required.  

Deletes numbers/settings.  

Takes user back up through the programming screens.

Allows set to be primed via the pump.

During an infusion repeatedly press to access infusion  

information: volume infused/to be infused, battery 

status, tube temperature and pumping pressure.

A2039-PocketGuide BG323 TPNMode AW.indd   1

1/12/10   16:20:36

323 Ambulatory Infusion Pump 

TPN Mode 

Quick user guide

Pump Alerts, Alarms and Trouble Shooting 

When an Alert Activates: 

Infusion continues, 3 beeps are heard. A message on the display 

screen alternates with the infusion running screen until the end  

of the battery or infusion. 

Low/end battery

 – the battery has approximately 30 minutes power 

remaining. Connect pump to charger or change battery pack. 

Program nearly complete

 – the programme/infusion has  

approximately 30 minutes until complete. Prepare to change  

bag and re-programme or close down the device. 

When an Alarm Activates: 

Infusion stops. LED indicator light turns red and the alarm sounds 

continuously until either the pump is paused or the problem is 

rectified. 

Always

 note the condition before pressing the ‘STOP/NO’ 

key to silence the alarm. 

Air/upstream occlusion

 – activates when either the upstream 

segment of the tubing is trapped/clamped or when more than 

the pre-set allowed volume of air has passed through the pump. 

Silence the alarm, check the upstream tubing is not occluded. 

Down occlusion

 – indicates that an occlusion has blocked the line 

or access device causing pressure build up in the set. The clamp 

may be closed or set trapped/kinked, which you should resolve. If 

successful the pump automatically resumes infusion. If you cannot 

detect and resolve the cause the access device may be blocked. 

Silence the alarm and contact your Clinician immediately. 

Refer over the page for manual and pump priming instructions. If re-

priming is indicated, ensure the line is disconnected from the patient. 

Door open

 – pump door has been opened during infusing. Simply 

close the door, silence the alarm and re-start. 

Pump unattended

 – pump has been left ‘stopped’ or unattended 

(may have programmed but forgotten to press Start). Alarm 

sounds after 2 minutes to indicate that the pump is not infusing. 

Restart pump

 – a system malfunction has occurred. Press the ‘INFO’  

key and record the error code. Some errors can be resolved by turning  

the pump off and on again. If when the pump powers on the message  

is displayed again, discontinue using the equipment, contact your 

Clinician/Provider immediately and report the error code. 

End program

 – alerts the user that the programme/infusion is  

now complete. 

Missing key

 – the administration set is loaded incorrectly.  

Check that the key is positioned correctly inside the pump  

before continuing. 

Wrong TPN Data

 – the programme entered by the user cannot be 

performed as it conflicts with either the operating parameters of the  

pump or the clinician/provider configurable settings (which require 

coded user access). Consult your clinician/provider for advice. 

Clinical contact: 
Phone number: 
E-mail contact: 

k  

The information contained in this guide is a summary only. 

Refer to Operating Manual for full operating instructions. 

k  

Screen information and sequences displayed are based on 

pump default settings and numerical values shown are 

examples only. You must refer to local policy/procedure  

for specific guidance on pump settings/set-up, protocols  

and use of accessories (e.g. cannula and administration lines). 

k  

When using the pump always follow screen prompts and 

before pressing keys to proceed ensure selections made 

correspond with what is required. 

k  

The information contained in this guide is applicable to 

software versions McTPN04, McTPN04C, McTPN05 and 

McTPN06. 

    

NOTE: With previous software versions some text wording/

screen information may vary slightly

.

CME

 

McKinley UK Ltd

 

 

Kincraig Business Park,  

Kincraig Road,  

Blackpool  FY2 0PJ 

Tel: 01253 894646  

Fax: 01253 896648  

email: [email protected]  

www.cme-mckinley.co.uk

CME McKinle

y/

323

/T

PN quick guide

/v2

/Dec 2010

CME McKinley Device Training is RCN Accredited 

The Multi-Function Keypad 

Keys ‘0’ to ‘9’ are used to enter numeric values during programming. 
Keys ‘2’ & ‘0’ are also used to scroll up and down respectively in menus. 

Powers the pump on and off. Press and hold until 

confirmation beep sounds to power off.

Starts/restarts the infusion. 

Confirms programming or selection during set up.

Stops the infusion when required.  

Deletes numbers/settings.  

Takes user back up through the programming screens.

Allows set to be primed via the pump.

During an infusion repeatedly press to access infusion  

information: volume infused/to be infused, battery 

status, tube temperature and pumping pressure.

A2039-PocketGuide BG323 TPNMode AW.indd   1

1/12/10   16:20:36

Do

c R

ef

: C

M

E M

ed

ic

al

/3

23

/T

PN Q

UG

/v

4/

Se

pt 2

01

4 - 

QUG0

9

 

CME Medical Device Training is RCN Accredited

323 Ambulatory Infusion Pump 

TPN Mode 

Quick user guide

Pump Alerts, Alarms and Trouble Shooting 

When an Alert Activates: 

Infusion continues, 3 beeps are heard. A message on the display 

screen alternates with the infusion running screen until the end  

of the battery or infusion. 

Low/end battery

 – the battery has approximately 30 minutes power 

remaining. Connect pump to charger or change battery pack. 

Program nearly complete

 – the programme/infusion has  

approximately 30 minutes until complete. Prepare to change  

bag and re-programme or close down the device. 

When an Alarm Activates: 

Infusion stops. LED indicator light turns red and the alarm sounds 

continuously until either the pump is paused or the problem is 

rectified. 

Always

 note the condition before pressing the ‘STOP/NO’ 

key to silence the alarm. 

Air/upstream occlusion

 – activates when either the upstream 

segment of the tubing is trapped/clamped or when more than 

the pre-set allowed volume of air has passed through the pump. 

Silence the alarm, check the upstream tubing is not occluded. 

Down occlusion

 – indicates that an occlusion has blocked the line 

or access device causing pressure build up in the set. The clamp 

may be closed or set trapped/kinked, which you should resolve. If 

successful the pump automatically resumes infusion. If you cannot 

detect and resolve the cause the access device may be blocked. 

Silence the alarm and contact your Clinician immediately. 

Refer over the page for manual and pump priming instructions. If re-

priming is indicated, ensure the line is disconnected from the patient. 

Door open

 – pump door has been opened during infusing. Simply 

close the door, silence the alarm and re-start. 

Pump unattended

 – pump has been left ‘stopped’ or unattended 

(may have programmed but forgotten to press Start). Alarm 

sounds after 2 minutes to indicate that the pump is not infusing. 

Restart pump

 – a system malfunction has occurred. Press the ‘INFO’  

key and record the error code. Some errors can be resolved by turning  

the pump off and on again. If when the pump powers on the message  

is displayed again, discontinue using the equipment, contact your 

Clinician/Provider immediately and report the error code. 

End program

 – alerts the user that the programme/infusion is  

now complete. 

Missing key

 – the administration set is loaded incorrectly.  

Check that the key is positioned correctly inside the pump  

before continuing. 

Wrong TPN Data

 – the programme entered by the user cannot be 

performed as it conflicts with either the operating parameters of the  

pump or the clinician/provider configurable settings (which require 

coded user access). Consult your clinician/provider for advice. 

Clinical contact: 
Phone number: 
E-mail contact: 

k  

The information contained in this guide is a summary only. 

Refer to Operating Manual for full operating instructions. 

k  

Screen information and sequences displayed are based on 

pump default settings and numerical values shown are 

examples only. You must refer to local policy/procedure  

for specific guidance on pump settings/set-up, protocols  

and use of accessories (e.g. cannula and administration lines). 

k  

When using the pump always follow screen prompts and 

before pressing keys to proceed ensure selections made 

correspond with what is required. 

k  

The information contained in this guide is applicable to 

software versions McTPN04, McTPN04C, McTPN05 and 

McTPN06. 

    

NOTE: With previous software versions some text wording/

screen information may vary slightly

.

CME

 

McKinley UK Ltd

 

 

Kincraig Business Park,  

Kincraig Road,  

Blackpool  FY2 0PJ 

Tel: 01253 894646  

Fax: 01253 896648  

email: [email protected]  

www.cme-mckinley.co.uk

CME McKinle

y/

323

/T

PN quick guide

/v2

/Dec 2010

CME McKinley Device Training is RCN Accredited 

The Multi-Function Keypad 

Keys ‘0’ to ‘9’ are used to enter numeric values during programming. 
Keys ‘2’ & ‘0’ are also used to scroll up and down respectively in menus. 

Powers the pump on and off. Press and hold until 

confirmation beep sounds to power off.

Starts/restarts the infusion. 

Confirms programming or selection during set up.

Stops the infusion when required.  

Deletes numbers/settings.  

Takes user back up through the programming screens.

Allows set to be primed via the pump.

During an infusion repeatedly press to access infusion  

information: volume infused/to be infused, battery 

status, tube temperature and pumping pressure.

A2039-PocketGuide BG323 TPNMode AW.indd   1

1/12/10   16:20:36

Kincraig Business Park, Kincraig Road, Blackpool FY2 0PJ  

Tel: 01253 894646 Fax: 01253 896648 [email protected]

323 Ambulatory Infusion Pump 

TPN Mode 

Quick user guide

Pump Alerts, Alarms and Trouble Shooting 

When an Alert Activates: 

Infusion continues, 3 beeps are heard. A message on the display 

screen alternates with the infusion running screen until the end  

of the battery or infusion. 

Low/end battery

 – the battery has approximately 30 minutes power 

remaining. Connect pump to charger or change battery pack. 

Program nearly complete

 – the programme/infusion has  

approximately 30 minutes until complete. Prepare to change  

bag and re-programme or close down the device. 

When an Alarm Activates: 

Infusion stops. LED indicator light turns red and the alarm sounds 

continuously until either the pump is paused or the problem is 

rectified. 

Always

 note the condition before pressing the ‘STOP/NO’ 

key to silence the alarm. 

Air/upstream occlusion

 – activates when either the upstream 

segment of the tubing is trapped/clamped or when more than 

the pre-set allowed volume of air has passed through the pump. 

Silence the alarm, check the upstream tubing is not occluded. 

Down occlusion

 – indicates that an occlusion has blocked the line 

or access device causing pressure build up in the set. The clamp 

may be closed or set trapped/kinked, which you should resolve. If 

successful the pump automatically resumes infusion. If you cannot 

detect and resolve the cause the access device may be blocked. 

Silence the alarm and contact your Clinician immediately. 

Refer over the page for manual and pump priming instructions. If re-

priming is indicated, ensure the line is disconnected from the patient. 

Door open

 – pump door has been opened during infusing. Simply 

close the door, silence the alarm and re-start. 

Pump unattended

 – pump has been left ‘stopped’ or unattended 

(may have programmed but forgotten to press Start). Alarm 

sounds after 2 minutes to indicate that the pump is not infusing. 

Restart pump

 – a system malfunction has occurred. Press the ‘INFO’  

key and record the error code. Some errors can be resolved by turning  

the pump off and on again. If when the pump powers on the message  

is displayed again, discontinue using the equipment, contact your 

Clinician/Provider immediately and report the error code. 

End program

 – alerts the user that the programme/infusion is  

now complete. 

Missing key

 – the administration set is loaded incorrectly.  

Check that the key is positioned correctly inside the pump  

before continuing. 

Wrong TPN Data

 – the programme entered by the user cannot be 

performed as it conflicts with either the operating parameters of the  

pump or the clinician/provider configurable settings (which require 

coded user access). Consult your clinician/provider for advice. 

Clinical contact: 
Phone number: 
E-mail contact: 

k  

The information contained in this guide is a summary only. 

Refer to Operating Manual for full operating instructions. 

k  

Screen information and sequences displayed are based on 

pump default settings and numerical values shown are 

examples only. You must refer to local policy/procedure  

for specific guidance on pump settings/set-up, protocols  

and use of accessories (e.g. cannula and administration lines). 

k  

When using the pump always follow screen prompts and 

before pressing keys to proceed ensure selections made 

correspond with what is required. 

k  

The information contained in this guide is applicable to 

software versions McTPN04, McTPN04C, McTPN05 and 

McTPN06. 

    

NOTE: With previous software versions some text wording/

screen information may vary slightly

.

CME

 

McKinley UK Ltd

 

 

Kincraig Business Park,  

Kincraig Road,  

Blackpool  FY2 0PJ 

Tel: 01253 894646  

Fax: 01253 896648  

email: [email protected]  

www.cme-mckinley.co.uk

CME McKinle

y/

323

/T

PN quick guide

/v2

/Dec 2010

CME McKinley Device Training is RCN Accredited 

The Multi-Function Keypad 

Keys ‘0’ to ‘9’ are used to enter numeric values during programming. 
Keys ‘2’ & ‘0’ are also used to scroll up and down respectively in menus. 

Powers the pump on and off. Press and hold until 

confirmation beep sounds to power off.

Starts/restarts the infusion. 

Confirms programming or selection during set up.

Stops the infusion when required.  

Deletes numbers/settings.  

Takes user back up through the programming screens.

Allows set to be primed via the pump.

During an infusion repeatedly press to access infusion  

information: volume infused/to be infused, battery 

status, tube temperature and pumping pressure.

A2039-PocketGuide BG323 TPNMode AW.indd   1

1/12/10   16:20:36

CME Medical Device and Advanced User training is RCN accredited 

and Skills for Health Quality Marked on all the guides.

The Multi-Function Keypad

Keys ‘0’ to ‘9’ are used to enter numeric values during programming. 
Keys ‘2’ & ‘0’ are also used to scroll up and down respectively in menus.

Reviews: