Cliq AS-1001D Instructions For Use Manual Download Page 1

AS-1001D

Suction Pump

Instructions for Use

 

Please read this manual carefully before using this device. Keep this manual for future reference.

  Cliq_AS-1001D Manual  (230V50Hz_20171124.3)

3510406D

Summary of Contents for AS-1001D

Page 1: ...AS 1001D Suction Pump Instructions for Use Please read this manual carefully before using this device Keep this manual for future reference Cliq_AS 1001D Manual 230V50Hz_20171124 3 3510406D...

Page 2: ...nt Parts Operation Cleaning Maintenance Repair Warranty Troubleshooting Transportation Storage and Disposal Guidance and Manufacturer s Declaration P 3 P 4 P 5 P 5 P 6 P 7 P 8 P 8 P 8 P 9 P 9 P 10 11...

Page 3: ...ins electrical and or electronic equipment that must be recycled and should be handed over to a suitable collection point for proper disposal Class II equipment Protection against electrical shock typ...

Page 4: ...I environment Further the device shall not be placed or operated in the following environments or near the following devices CT diathermy machine RFID transmitters electromagnetic security systems suc...

Page 5: ...atheters that are compliant with local and or national laws regulations and directives In the EU only use suction catheters that are labeled with the CE mark are MDD compliant and have received market...

Page 6: ...12 WARNING Suction catheters are not provided with this device and must be purchased separately in accordance with the instructions of a trained healthcare professional or physician Only use disposabl...

Page 7: ...suction tubing remains completely blocked the vacuum level should remain steady For suctioning of an adult airway experts recommend a vacuum level between 100 150 mmHg Once unit is set to the desired...

Page 8: ...ories such as the suction catheter must be disposed of according to local regulations institution policy as applicable and the instructions of a certified healthcare professional Wear disposable glove...

Page 9: ...rdous Substances Directive prohibits the disposal of this device in the domestic waste Outside the EU Follow disposal regulations in the country of use Problem Possible cause Remedy Unit does not star...

Page 10: ...S 1001D suction pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the AS 1001D suction pump can help prevent elect...

Page 11: ...commercial or hospital environment If the user of the EQUIPMENT or SYSTEM requires continued operation during power mains interruptions it is recommended that the EQUIPMENT or SYSTEM be powered from...

Page 12: ...wan 11011 www cliqmedical com info cliqmedical com Tel 886 2 2345 1868 4B 03 Taipei World Trade Center 4F No 5 Sec 5 HsinYi Rd Taipei Taiwan 11011 Authorized Service Center Contact Infomation Copyrigh...