CLEAN Instruments PH500T Operation Quick Manual Download Page 1

PH500T pH Meter Operation Quick Guide

 

 

PH500T pH Meter Operation 
Quick Guide 

1.

 

Specification 

Parameters: pH, mV, Temperature 
pH level: 0.001pH 
pH range: (-2.000~20.000)pH 
mV range: (-2000.00~2000.00)mV 
Temperature range: (-10.0~135.0)℃

/(14.0  ~  275.0) ℉ 

2.

 

Screen Icons 

 

 2.1.

 

Annotation   

No. 

Explanation 

User ID   

Measurement parameters 

Reading states 

Measurement information 

Method management 

Tips information 

System time   

Calibration information 

No. 

Explanation 

Sample ID     

10 

Soft function buttons     

 2.2.

 

Symbol 

Symbol 

Explanation 

 

Reading status, display the 

measurement status of reading, stable, 

locked each indicates that the 

processing, stable, and reading 

completed. 

PTS 

The percentage slope of the pH 

electrode calibration data 

ATC 

Auto Temperature compensation 

MTC 

Manual temperature compensation 

ORP 

Redox potential value, in mV 

Offset 

Offset potential, in mV 

 

Measurement method management, 

display the current method information 

 

Standard solution for pH electrode 

calibration 

 

User ID 

 

Sample ID 

3.

 

pH Operation Quick Guide 

 3.1.

 

Preparation   

1.

 

Install the components of the 
instrument and connect the pH and 
temperature electrode. 

2.

 

Prepare standard buffer solutions 
such as pH4.01, pH7.00, pH10.01 
standard solutions. 

3.

 

Remove the protective cap at the 
lower end of the pH electrode, pull 
down the rubber cover at the upper 
end of the electrode. Expose the top 
hole, rinse the electrode with distilled 
water, and dry with filter paper. 

4.

 

Press the power key to turn on the 
instrument. 

5.

 

Select measurement method: 

●  The default measurement method of 
the instrument is "No.001 direct reading 
measurement method", pH measurement 
parameter, continuous reading mode, if 
adopting the default method for 
measurement, this step can be omitted. 

●  Press 

to enter the measurement   

method management. You can select the 
method in the method list or create new 
methods as required. 

6.      The main measurement parameter 
settings of the instrument: 

Reviews: