U-PAD2 A/D Converter
4 Using the U-PAD2
4.2.2 Method Setup - Measurement
Fig 17: Method Setup - Measurement
External Start/Stop
Enables control from an external signal.
Note:
The input used for an external start from a chromatograph can be set in
the
dialog, see
Start Only
The external signal only starts an analysis. Triggering the input will not
influence the acquisition in progress.
Start - Restart
The external signal terminates the analysis in progress and at the
same time starts a new one. The first signal starts the analysis; the
second signal terminates it and at the same time starts a new one - the
so-called continuous measurement.
Start - Stop
The external signal starts and stops an analysis. The first signal starts
the analysis; the second signal stops it, etc.
Down
The station reacts to changes in voltage in relevant controlling input
from high (> 3 V) to low (< 0.7 V), or to contacts closed in the relay.
Up
The station reacts to changes in voltage in relevant controlling input
from low (< 0.7 V) to high (> 3 V), or to contacts opened in the relay.
- 18 -