Maintenance and Support
153
RetCam 3 and RetCam Shuttle User Manual
PN 18-000393 Rev. A
If you clean or disinfect, rinse the lens with sterile water to avoid corneal
de-epithelialization that can be caused by residual solution.
Since disinfection solutions can damage the lens piece, do not exceed the
recommended soak times.
Alternate Cleaning and Disinfection
• Cleaning Solution (Double): FrankLab Enzymex L9 (immersion time 5 to 15 minutes;
no longer than 2 hours)
• Disinfection: FrankLab Peraxy-Lens or Peralex H8+ (30 minutes minimum; no longer
than 2 hours)
Please refer to the specific FrankLab instruction on double cleaning and disinfection.
Figure 141
Inverted Lens Piece Showing Permissible Immersion Depth
Cleaning the Rest of the System
As with any medical device, use good public health practices when handling the
equipment, based on CDC guidelines. In addition, as with typical office equipment, a
gentle wiping with a cloth moistened with mild soap and/or water is recommended. Do not
spray cleaning solutions on any part of the system.
Disposal of Materials
Dispose of waste materials according to local and national requirements. Contact Clarity
Technical Services if additional assistance is required.