CDB5480U
DS893DB5
25
Figure 20.
Calibration
Window
The
Refresh Screen
button will update the contents of the screen by reading all the register values from
the part. It is a good idea to press the
Refresh Screen
button when entering the
Calibration
window, or
after modifying any registers to reflect the current status of the CS5480.
2.6.1
Save Cal and Load Cal Buttons
Clicking the
Save Cal
button will save the calibration widow's configuration to a .txt file. Clicking the
Load
Cal
button will recall a saved configuration and store values into the CS5480.
2.6.2
Offset / Gain Register
In the
Offset
and
Gain Calibration
boxes, the offset and gain registers for all channels are displayed in
hexadecimal and decimal formats. These registers can be modified directly by typing the desired value in
the display boxes. There are three types of offset registers: DC offset, AC offset and power offset. The
AC offset registers only affect the RMS register values. The active and reactive power offset registers only
affect the active and reactive power register values, respectively. The DC, AC, and power offset registers
are two's complement numbers whose values range from -1 to +1. The gain register value ranges from 0
to 4.