OPERATING INSTRUCTIONS 

DLM-260 

DIGITAL LCR METER 

 

 

SPECIFICATIONS 

Display: 

3

½

  digit  liquid  crystal  display  (LCD)  with  a 

maximum reading of 1999. 

Polarity: 

Automatic,  positive  implied,  negative  polarity 

indication. 

Overrange

: (OL) or (-OL) is displayed. 

Zero: 

Automatic. 

Low battery indication: 

The “

” is displayed when the 

battery voltage drops below the operating level. 

Measurement rate: 

2.5 times per second, nominal. 

Operating Environment: 

0°C to 40°C at <70% R.H. 

Storage  Temperature: 

-20°C  to  60°C,  0  to  80%  R.H. 

with battery removed from meter. 

Accuracy: 

Stated accuracy at 23°C±5°C, <75% R.H. 

Power: 

Single  standard  9-volt  battery,  NEDA  1604,  JIS 

006P, IEC 6F22. 

Battery life: 

60 hours typical with carbon-zinc. 

Dimensions: 

200mm (H) x 90mm (W) x 40mm (D). 

Weight: 

Approx. 14 oz. (400g) including battery. 

Accessories: 

One pair test leads, one spare fuse installed, 

9V battery (installed) and Operating Instructions. 

 

DC VOLTS 

Ranges: 

20V 

Accuracy: 

±(2.0%rdg + 1dgt) 

Input impedance: 

1M

Ω

 

Overload protection: 

25V DC or AC rms 

 

TEMPERATURE 

Ranges: 

-20°C to 750°C 

Resolution: 

1°C 

Accuracy:   

±(2.0%rdg + 3dgts) on -20°C to 500°C 
±(3.0%rdg + 2dgts) on 500°C to 750°C 
 

DIODE TEST 

Including: 

, continuity test 

Test current: 

1.0mA±0.6mA 

Open voltage: 

3.0VDC typical 

Accuracy: 

±(3.0%rdg +1dgt) 

Continuity: 

<30dgts 

Display: 

Forward junction voltage 

Overload protection: 

25V DC or AC rms 

 

SIGNAL OUTPUT 

Signal: 

+3V, -0.5V square wave, 50% duty 

Voltage:    

Hi: +5V approx 

Lo: -2V approx 

Frequency: 

2.5KHz square wave 

Output impedance: 

3.5K

Ω

 

Overload protection: 

25V DC or AC rms 

 

FREQUENCY 

(Autoranging) 

Ranges: 

2KHz, 20KHz, 200KHz, 2000KHz, 15MHz 

Accuracy: 

±(0.1%rdg +1dgt) 

Sensitivity: 

1.0Vrms min (TTL signal) 

Overload protection: 

25V DC or AC rms 

 

CAPACITANCE 

Ranges: 

200pF,  2nF,  20nF,  200nF,  2µF,  20µF,  200µF, 

2000µF, 20mF 

Accuracy: 

±(2.0%rdg + 30dgts) on 200pF range 
±(2.0%rdg + 10dgts) on 2nF to 20µF ranges 
±(3.0%rdg + 10dgts) on 200µF to 20mF ranges 

Test frequency: 

1000Hz on 200pF to 20nF ranges 
80Hz on 200nF to 2µF ranges 
26Hz on 20µF range 
10.5Hz on 200µF to 20mF ranges 

Overload protection: 

0.1A/250V fast blow fuse 

Note: In lower range 200pF, 2nF subtract residual offset 

reading from result with test leads opening.

 

 

RESISTANCE 

Ranges: 

200

Ω

,  2K

Ω

,  20K

Ω

,  200K

Ω

,  2000K

Ω

,  20M

Ω

200M

Ω

, 2000M

Ω

 

Resolution: 

200

Ω

 range 100m

Ω

 

Accuracy: 

±(0.3%rdg + 3dgts) on 200

Ω

 range 

±(0.3%rdg + 1dgt) on 2K

Ω

 to 2000K

Ω

 ranges 

±(2.0%rdg + 2dgts) on 20M

Ω

 range 

±[(5.0%rdg - 10dgts) + 10dgts] on 200M

Ω

 to 2000M

Ω

 

ranges 

Open circuit volts: 

3.0VDC on 200

Ω

, 200M

Ω

, 2000M

Ω

 ranges 

0.3VDC on other ranges 

Overload protection: 

25V DC or AC rms 

 

INDUCTANCE 

Ranges: 

200µH, 2mH, 20mH, 200mH, 2H, 20H 

Accuracy: 

±(5.0%rdg + 3dgts) 

Test frequency: 

1000Hz on 200µH to 20mH ranges 
80Hz on 200mH to 2H ranges 

26Hz on 20H range 

Overload protection: 

0.1A/250V fast blow fuse 

Note:  In  lower  range  200µH,  2mH  subtract  residual 

offset  reading  from  result  with  test  leads  being 
shorted. 

OPERATION 

However,  electrical  noise  or  intense  electromagnetic 

fields  in  the  equipment  may  disturb  the  measurement  cir-
cuit. Measuring instruments will also respond to unwanted 
signals that may be present within the measurement circuit. 
Users  should  exercise  care  and  take  appropriate  precau-
tions  to  avoid  misleading  results  when  making  measure-
ments in the presence of electronic interference. 

 

Voltage Measurements 

1.  Connect  the  red  test  lead  to  the  DCV-DIODE-Hz  “

+

” 

jack and the black test lead to the DCV-DIODE-Hz “

” 

jack. 

2. Set the Function/Range switch to the DC 20V range. 
3.  Connect  the  test  leads  to  the  device  or  circuit  being 

measured. 

4.  For  dc,  a  (

)  sign  is  displayed  for  negative  polarity, 

positive polarity is implied. 
 

Temperature Measurements 

WARNING 

Remove test leads being measured.

 

1. Set the Function/Range switch to the “°C” position. 
2.  Connect  a  type  k  thermocouple  to  the  jack  on  the  in-

strument.  Place  the  probe  or  thermocouple  tip  on  or  in 
the  material  to  be  measured  and  take  the  temperature 
reading directly from the display. 

 

Diode Tests and Continuity Measurements 

1.  Connect  the  red  test  lead  to  the  DCV-DIODE-Hz  “+” 

jack and the black test lead to the DCV-DIODE-Hz “

” 

jack. 

2. Set the Function/Range switch to the 

  position. 

3. Turn off power to the circuit under test. 
4.  Touch  probes  to  diodes.  A  forward-voltage  drop  on 

diode. Microwave diode about 0.6VDC typical. 

5.  Reverse  probes.  If  the  diode  is  good,  display  rending 

“OL”. 

6. If the junction is measured in a circuit and a low reading 

is obtained with both lead connections, the junction may 
be shunted by a resistance of less than 1k

Ω

. In this case 

the diode must be disconnected from the circuit for ac-
curate testing. 

7.  If  display  reading  <30dgts,  the  beeper  sounds  continu-

ously. 

 

Signal Output 

1. Set the Function/Range switch to the “

” position. 

2.  Connect  the  red  test  lead  to  the  DCV-DIODE-Hz  “+” 

jack and the black test lead to the DCV-DIODE-Hz “

” 

jack. 

3. Connect the test leads to the points of signal input. 
 

 