background image

 

MetCam Technical Manual

 

Document P/N A1604870010 

14 

 

 

4.6

Commissioning 

For commissioning, refer to "MetCam Commissioning and WebUI Manual" document. 

 

5

Troubleshooting 

Problem 

 

Cause

 

Corrective action

 

After power-up, the LED does not light 
up at all 
4-20 mA outputs 0 mA value 

Power connection polarity 
is reversed 

 

 

Check power connection polarity 

 

If polarity is correct and power is live, send 
MetCam to CI for repairs 

After power-up, the LED is a constant 
red ("ERR" status) 

Power is not within 
allowed range 

 

Check power voltage level. 

 

Reset the power and allow MetCam to try 
and boot again 

 

If Level is correct, send MetCam to CI for 
repairs 

After power-up, the Green LED does 
not stop blinking and does not turn into 
a constant Green (stays in the "BOOT" 
status) 

Internal MetCam problem 

 

Reset the power and allow MetCam to try 
and boot again 

 

If the problem persists, send MetCam to CI 
for repairs 

 

6

Maintenance 

Maintenance should be performed by suitably qualified personnel who are familiar with local codes and practice. MetCam 
must be kept as clean as possible. Clean the MetCam viewing windows and the reflector periodically.  

The frequency of cleaning operations depends upon the local environmental conditions and specific applications. It is 
recommended to visually inspect (and clean if deemed necessary) once quarterly.   
Warnings 2 and 3 (see list in ) could indicate that an optical window cleaning is due. 

It is highly probable that MetCam will alarm during the cleaning. There is no need to cut off the power to MetCam prior to 
cleaning.  

NOTICE 

Do not use solvents or abrasives for cleaning MetCam optics. Failure to comply could result in damage to 

MetCam optics and/or its surfaces. 

 

6.1

Optical Cleaning Instructions 

In  case  an  optical  surface  has  been  contaminated,  clean  using  only  wipes  approved  for  optical  equipment  with 

Isopropyl Alcohol

. Finish with clean optical tissue. DO NOT TOUCH the optical surfaces with bare hands

.

 

 

6.2

Exterior Cleaning Instructions 

Where dust, dirt, or moisture accumulates on MetCam body, wipe clean with a soft cloth and detergent only and 
then rinse with clean water. 

NOTICE 

If force was applied during the cleaning, it is advised to check if the MetCam alignment is still good. Small 

movements do not have an impact as MetCam covers a large field of view.  

It could, however, be necessary to verify the location of the ROI (See "MetCam commissioning and WebUI" document) 

Summary of Contents for MetCam A160100 0 Series

Page 1: ...MetCam Technical Manual Document P N A1604870010 1 MetCam Models A160100xxx0 Instructions for Use Manual Rev B Feb 2020...

Page 2: ...Type Marking 6 4 Use 7 4 1 Installation Prerequisites 7 4 1 1 Electrical Installation Prerequisites 7 4 2 Installation 8 4 2 1 Handling 8 4 2 2 Installing the Tilt Mount 9 4 2 3 Connecting MetCam to t...

Page 3: ...ocument P N A1604870010 3 5 Troubleshooting 14 6 Maintenance 14 6 1 Optical Cleaning Instructions 14 6 2 Exterior Cleaning Instructions 14 6 3 Periodic Procedures 15 7 Disposal 15 8 Technical Data 15...

Page 4: ...ational rules and regulations associated with this product Only trained and competent personnel who are familiar with local codes and practice are permitted to inspect repair and service the product a...

Page 5: ...d The threaded back plate shall be tightened all the way to the stopping shoulder and secured against self loosening by tightening the set screw using hex 0 050 Lens guards must always be applied and...

Page 6: ...as a supplementary detection device for total risk reduction The limitations of this technology must be assessed with respect to their ability or inability to determine the concentration of gas There...

Page 7: ...here is a significant likelihood of equipment personnel or moving objects accidentally knocking MetCam out of alignment should be avoided If such locations cannot be avoided measures including improve...

Page 8: ...and wire MetCam see Section 4 3 1 MetCam power up see Section 4 4 4 2 1 Handling Warranty void if opening MetCam enclosure other than the back cover used for accessing the electrical connections to th...

Page 9: ...mount is not symmetrical make sure to install it in the correct orientation Note that the tilt mount does not include a pan degree of freedom so it is important to align it to the monitored area when...

Page 10: ...eld optional item should be mounted after connecting MetCam to the tilt mount It does not interfere with wiring or with aiming MetCam 4 2 5 Installing the Wi Fi Antenna MetCam can be ordered with an o...

Page 11: ...the back of MetCam needs to be opened for accessing the terminal block and then needs to be closed properly so that the unit will retain its explosion flame proof certification The arrow in the pictur...

Page 12: ...vailable for connecting the cable shield on the cable gland entrance collar Terminal Block Pin Assignment Pin Marking on Terminal Block Description 1 Voltage input Max 32 V 2 Voltage input 3 4 20 mA H...

Page 13: ...cation of the Indicator LED the green arrow is pointing to it 4 5 4 20 mA Signaling and LED Status MetCam uses a stepped output and needs to be installed as a 4 wire transmitter The signaling range is...

Page 14: ...reflector periodically The frequency of cleaning operations depends upon the local environmental conditions and specific applications It is recommended to visually inspect and clean if deemed necessa...

Page 15: ...t be disposed of as household waste You can return this product to CI Systems free of charge For information please contact the national marketing organizations or CI Systems 8 Technical Data Mechanic...

Page 16: ...temperature range w o wifi A1601001340 Extended temperature range wifi M25 Gland Entrance MetCam Models P N Description A1601000200 Standard temperature range w o wifi A1601000240 Standard temperature...

Reviews: