CD40047_R04_allPIXA_User-Manual.docx
186
Tracing
For test and debugging purposes, the camera writes process data to an internal logging
memory. Which data are logged, can be selected by selecting check boxes on the
Tracing
page.
For more information, ask the Chromasens support.
E-Mail:
NOTE
Logging of process data reduces performance of the internal
microcontroller. Therefore the internal trace should only be activated for
testing. All tracing should be deactivated in real production mode of the
camera.
11.16 Line trigger and encoder setup (Line synchronisation)
NOTE I
Due to internal timing limitations there is a small delay in the start of the
integration time after the line trigger and this delay might jitter. Therefore, a
pulsed illumination which is synchronous to the trigger signal might cause
problems in image quality.
NOTE II
If
the camera is used in the “Triggered Frame Scan” with active LineTrigger or
encoder, the linesync (Encoder or LineTrigger) has to fit some pulses before light
barrier input.
Enable encoder
Name
Name
Enable encoder
Function
Enables the line trigger or encoder mode in the camera and the main selection
between internal and external line trigger.
Without encoder, camera speed is defined by the values of
Integration time
and optionally by
Line period time
. If encoder is enabled, camera speed if
defined by an external line trigger or by encoder pulses.
Channels
All
Unit
Bool
Dependency
None
Notes
If you use this mode, the external ports must be connected with the internal
function using the IO Configurator of CST.
Camera
Function
Lower limit
All
Off / Internal
Upper limit
All
On / external