Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
5-26
Table 5-8
Bit No. Bit Weight
Description
7
128
Power on Bit. Reboot the Power Supply can set
this bit to 1.
6
Always 0.
5
32
Command Error Bit. This bit is set to 1 if there is
any IEEE 488.2 syntax error.
4
16
Execution Error Bit. This bit is set to 1 when the
command parameter is out of valid range or
inconsistent.
3
8
Device Dependent Error Bit. This bit is set to 1
when too many errors have occurred that the
error queue is full.
2
4
Queue Error Bit. This bit is set to 1 when reading
data from the output buffer and no data is present,
or when the data is lost.
1
Always 0.
0
1
5.6.2
Specific Commands for 62000H Series
5.6.2.1
ABORT Subsystem
ABORt
Description:
It sets all output state to “OFF”.
Syntax:
ABORt
5.6.2.2
CONFIGURE Subsystem
1.
CONFigure:BEEPer
Description:
It sets the beeper to ON or OFF.
Syntax:
CONFigure:BEEPer ON
CONFigure:BEEPer OFF
Parameter:
ON|OFF
Example:
CONF: BEEPer ON
CONF: BEEPer OFF
Query Syntax:
CONFigure:BEEPer?
Return Parameter: ON | OFF
Query Example: CONF:BEEPer?
It returns the beeper control status.
Return Example: ON or OFF
2.
CONFigure:REMote
Description:
It sets the remote control status (valid for RS232C only).
Syntax:
CONFigure:REMote ON
CONFigure:REMote OFF
Parameter:
ON|OFF
Example:
CONF:REM OFF
It disables remote control.