![Chroma 61501 User Manual Download Page 21](http://html1.mh-extra.com/html/chroma/61501/61501_user-manual_2604431021.webp)
Installation
2-1
2. Installation
2.1 Inspection
After unpacking the instrument, please inspect any damage that may have occurred during
the shipment. Save all packing materials in case the instrument has to be returned one day.
If any damage is found, please file a claim with the carrier immediately. Do not return the
instrument to the factory without obtaining the prior RMA acceptance from Chroma.
2.2 Preparation for the Use
In the beginning, the instrument must be connected with an appropriate AC line input. Then,
since fans intelligently cool it, it must be installed in sufficient space for circulation of air. It
should be used in an area where the ambient temperature does not exceed 40
°
C.
2.3 Requirements of Input Power
2.3.1 Ratings
Input Voltage Range
:
90 ~ 250 Vac, single phase
Input Frequency
:
47-63 Hz
Max. Current
:
61501 : 8 A
61502
:
16
A
61503
:
21
A
61504 : 28 A
CAUTION
AC source may be damaged if it is operated at an input voltage that is
outside its configured input range.
2.3.2 Input Connection
The input terminal is located at the bottom of the instrument rear panel. The rated withstand
temperature of power cord must be at least 85
°
C or above. The current carried by power
line input should be greater than or equal to the maximum rated input current of AC source.
See Figure 2-1 and do the following things one by one:
1.
Remove the safety cover from the back of AC source.
2.
Plug in the power cord to the INPUT terminal of AC source as described below:
Green or green/yellow wire to the terminal labeled “G”.
White or blue wire to the terminal labeled “N”.
Black or brown wire to the terminal labeled “L”.
3.
Slide the safety cover over the AC input terminal, and secure the cover with two screws.
Summary of Contents for 61501
Page 1: ...i Programmable AC Source 61501 61502 61503 61504 User s Manual ...
Page 2: ......
Page 3: ...Programmable AC Source 61501 61502 61503 61504 User s Manual Version 1 6 January 2010 ...
Page 7: ...v ...
Page 20: ......
Page 26: ......
Page 28: ...Programmable AC Source 61501 61502 61503 61504 User s Manual 3 2 ...
Page 29: ...Local Operation 3 3 Figure 3 1 ...
Page 56: ......
Page 118: ......
Page 134: ......