Calibration Procedure
4-13
(5) Press [ENTER] for saving the calibration value.
(6) Press [STOP] to stop GC 5
Ω
Full calibration.
4.7
LCD Contrast Calibration
■
After cleared the memory data, this calibration set LCD Contrast to the best suitable
value.
(1) Press [UP] or [DOWN] to select LCD Contrast Calibration.
(2) Press [SETUP] to enter LCD Contrast setting.
(3) Press [INC.] or [DEC.] until the reading shows the best suitable brightness.
4.8
Complete Calibration
■
Press [EXIT] to complete the calibration. Press the front panel calibration switch again
for opening the hardware data backup protection circuit to avoid the calibration data loss.