Operation and Setting
2-9
2.3.3 PRESET Parameter Setting
■
How to enter the PRESET menu
In
[Standby Menu]
, press Function Key [PRESET] to enter the PRESET menu as shown
below:
UP
1.ACV FREQUENCY:
60Hz
2.SOFTWARE AGC :
ON
DOWN
3.WV AUTO-RANGE:
OFF
4.IR AUTO-RANGE:
ON
NEXT
5.GFI :
ON
EXIT
50Hz/60Hz
RMT LOCK OFST ERR
(1) After entering the PRESET parameter setting menu, press Function Key [F3] to
move the highlight to the parameter item you desired to set.
(2) Press Function Key [F1] and [F2] to set the parameter data.
(3) PRESET parameter function table as shown below:
Item
Range Default
Description
ACV FREQUENCY 50 / 60 Hz 60Hz
It sets the output voltage frequency at AC HiPot
test.
SOFTWARE AGC
OFF / ON
ON
It sets if to open the software AGC function.
WV AUTO-RANGE OFF / ON
OFF It sets if to open the HiPot auto range function.
IR AUTO-RANGE
OFF / ON
ON
It sets if to open the insulation auto range
function.
GFI
OFF / ON
ON
It sets if to open the GFI function.
FAIL RESTART
OFF / ON
OFF It sets if to restart directly when fails the test.
SCREEN
OFF/ON
ON
It sets if to display the test screen.
2.3.4 PROGRAM Parameter Setting
■
Test parameter setting
(1) In
[Standby Menu]
, press Function Key [PROGRAM] to enter the PROGRAM
setting menu as the following:
STEP
1/1
AC
LOW : OFF
NEW
ARC : OFF
VOLT :0.050kV
RAMP : OFF
MORE..
HIGH :0.500mA
FALL : OFF
TIME : 3.0s
NEXT
EXIT
1-10
RMT LOCK OFST ERR
■
Test steps setting
(1) Press [F3] to move the highlight and stay in the test item. When Function Key [F1]
shows “UP”, it means press this key can switch to the next test step. When Function