Description
Single-lamp mode
Dual-lamp mode
A
Distance above the floor
which the power
measurements must be
below 0.76uW/cm
2
2.1 m (7 feet)
2.1 m (7 feet)
B
Installation height of the
CounterAct device
6.7 m (22 feet) to 8.5 m (28 feet)
8.5 m (28 feet) or greater
Setting the lamp mode
CounterAct is equipped with two Care222 B1 lamp modules and can operate in single-lamp or dual-
lamp mode.
Caution! If not avoided, the following could result in minor or moderate injury.
•
This procedure must be performed by a qualified installer.
•
Single-lamp mode—Runs one lamp at a time when disinfecting and intelligently equalizes the
operating hours for the two lamp modules. CounterAct must be installed in a ceiling with a
minimum height of 6.7 m (22 feet) to 8.5 m (28 feet).
The estimated lamp replacement cycle for CounterAct operating in single-lamp mode is double
to the lamp replacement cycle operating in dual-lamp mode.
•
Dual-lamp mode—Runs both lamps when disinfecting. CounterAct must be installed in a ceiling
with a minimum height of 8.5 m (28 feet).
To ensure safety redundant circuitry exists, CounterAct has two lamp mode switches. The following
image shows the position of the switches (on the reverse side of the Care222 label).
Installation and setup
CounterAct Installation and Setup Guide
19
020-103456-01 Rev. 1 (09-2021)