Rotational Vacuum Concentrator RVC 2-33 CDplus
with infrared heating
8 Maintenance and service
82 / 105
Version 08/2009, Rev. 3.6 of 22/11/2021
• sb-bs
Translation of the original operating manual
8.3
Autoclaving
The service life of the accessories essentially depends on the frequency of
autoclaving and use.
• Remove the existing O-rings so that the trapezoid notch can be reached
by the hot steam. This prevents the formation of humidity in the notch..
• Replace the accessories immediately when the parts show changes in
colour or structure or in the occurrence of leaks etc.
Accessories
Max. temp
[°C]
Min. time
[min]
Max. time
[min]
Max. cycles
Glass tubes
134-138
3
40
-
Polycarbonate tubes
115-118
30
40
20
Polypropylene tubes
115-118
30
40
30
Teflon tubes
134-138
3
5
100
Aluminium rotors
134-138
3
20
-
Aluminium accessories
134-138
3
20
-
Pos : 204 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1
Pos : 205 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 080 W art ung und Inst andhalt ung/080- 0040 I nst andhalt ungsar beit en @ 25\m od_1404985246314_68.docx @ 187625 @ 2 @ 1
→
8.4
Service
DANGER
Danger to life due to electric shock!
In the event of service work that requires the removal of the panels, there
is a risk of electric shock or mechanical injury. Only qualified specialist
personnel is authorised to perform this service work.
The rotational vacuum concentrator is subject to high mechanical stress. In
order to be able to withstand this high level of stress, high-quality
components were used during the production of the rotational vacuum
concentrators. Nevertheless, wear cannot be excluded and it may not be
visible from the outside.
This is why we recommend having the rotational vacuum concentrator
checked by the manufacturer during an inspection once per year in the
operating state and once every three years in the dismantled state.
Information and appointments:
In Germany:
Contact
Martin Christ Gefriertrocknungsanlagen GmbH
An der Unteren Söse 50
37520 Osterode
Tel. +49 (0) 55 22 / 50 07-44 44
E-mail: [email protected]