5 Set-up and connection
Version 08/2009, Rev. 3.6 of 22/11/2021
• sb-bs
35 / 105
Translation of the original operating manual
Pos : 104 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 050 Aufst ellung und Anschl uss /050 Aufst ellung und Ans chl uss========================== @ 25\m od_1404985082878_68.docx @ 186981 @ 1 @ 1
→
5
Set-up and connection
Pos : 105 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 050 Aufst ellung und Anschl uss /050- 0010 Aufs tell ort, Einsatzort @ 25\m od_1404985083707_68. docx @ 186995 @ 2 @ 1
→
5.1
Installation site
Operate the RVC only in closed and dry rooms.
• The table must be stable and have a solid, even tabletop.
• Ensure sufficient ventilation. Do not place any paper, cloth, or similar
material behind or under the unit, since otherwise the air circulation will
be impaired.
• Keep a safety distance of at least 30 cm from the wall so that the vents
in the unit remain fully effective.
• The ambient temperature must be in the range of +10°C to +25°C.
• Do not subject the RVC to thermal stress, e.g. by positioning it near
heat generators.
• Avoid direct sunlight (UV radiation).
Pos : 106 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 107 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 050 Aufst ellung und Anschl uss /050- 0020 Energievers orgung--------------------------------------- @ 25\m od_1404985084585_68. docx @ 187009 @ 2 @ 1
→
5.2
Power supply
Pos : 108 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 050 Aufst ellung und Anschl uss /050- 0020-0010 Ans chl uss art (Kaltger ät est eck er) @ 8\m od_1309335834268_68.docx @ 46621 @ 3 @ 1
→
5.2.1
Type of connection
DANGER
Danger to life due to electric shock!
The operating voltage on the name plate must correspond to the local
supply voltage!
Christ rotational vacuum concentrators are units of protection class I.
Rotational vacuum concentrators of this type have a three-wire power cord
with an IEC C13 connector.
NOTE
The removable power cord must not be replaced with a power cord of
inadequate rating!
An equipotential bonding screw is located on the back (see chapter 10 -
"Technical data"), This equipotential bonding screw can be used to perform
an earth conductor check.
(see chapter 2.1.1 - "Functional and operating elements")
Pos : 109 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 110 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 050 Aufst ellung und Anschl uss /050- 0020-0020 Sicherungen baus eits @ 25\mod_1404985085466_68. docx @ 187023 @ 3 @ 1
→
5.2.2
Customer-provided fuses
Sufficiently rated fuse protection of the rotational vacuum concentrator in
the electrical system of the building is required.
Pos : 111 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1