7 Malfunctions and error correction
50 / 84
Version 04/2012, Rev. 2.3 of 03/07/2017
• sb
Translation of the original operating manual
7
Malfunctions and error correction
7.1
General malfunctions
Malfunctions are displayed in the menu "Process & equipment info" (see
chapter 6.5.3.2 - "Process and equipment information"). A sound signal
sounds when an error message is generated.
• Eliminate the source of the problem (see the following chapters).
• Acknowledge the error messages by pressing the right-hand function
key.
7.1.1
Error messages
Error
Possible reason
Correction
Lid cannot be opened
•
The rotor chamber is not
aerated completely
•
Aerate the rotor chamber
completely
Lid contacts error
•
The system has detected an
open lid during the evaporation.
The rotor will be stopped
immediately. The lid must be
closed while the rotor is moving.
•
Check the lid contacts. If
necessary, contact the after-
sales service (see chapter 7.2 -
"Service contact")
Rotor overtemperature
•
The maximum permissible rotor
temperature has been
exceeded
•
If the system does not cool
down, disconnect it immediately
from the power supply.
•
Contact the after-sales service
(see chapter 7.2 - "Service
contact")
Chamber overtemperature
•
The maximum permissible
chamber temperature has been
exceeded.
•
If the system does not cool
down, disconnect it immediately
from the power supply.
•
Contact the after-sales-service
(see chapter 7.2 - "Service
contact").
Permissible imbalance exceeded
•
The rotor will be stopped
immediately, the process will
also be stopped.
– improper loading of the
rotor
– glass breakage during the
evaporation
– uneven evaporation
•
Eliminate the imbalance and
restart the unit. If the error is still
present, contact the after sales-
service (see chapter 7.2 -
"Service contact")
VSK3000 calibration error
•
Incorrect calibration of the
vacuum sensor VSK3000
•
The sensor must be recalibrated
(see the separate operating
manual).