Freeze-dryer Alpha 1-4 LDplus
Freeze-dryer Alpha 2-4 LDplus
7 Malfunctions and error correction
Version 11/2006, Rev. 2.2 of 17/10/2016
• sb
53 / 102
Translation of the original operating manual
Pos : 143 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070 St örungen und F ehl ersuche=========================== @ 25\ mod_1404984056039_68. doc x @ 184111 @ 1 @ 1
7
Malfunctions and error correction
Malfunctions are displayed in the dialog box "Process & equipment
messages" (see chapter 6.5.3.3 - "Process and equipment information").
An acoustic signal sounds when an error message is generated.
• Eliminate the source of the problem (see the following chapter).
• Acknowledge the error message.
Pos : 144 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 145 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010 Allgemei ne Stör ung en LD plus @ 31\ mod_1408438256415_68.doc x @ 209704 @ 2 @ 1
7.1
General malfunctions
Pos : 146 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 147 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0010 Netzaus fall @ 25\ mod_1404984058964_68. doc x @ 184153 @ 3 @ 1
7.1.1
Power failure
The control system continues with the process after a power failure. The
preselected conditions remain saved even during a process run.
In the event of a power failure in the drying phase, the batch may become
unusable. Whether the batch can be saved or not depends on the drying
phase in which the product was when the power failure occurred.
• In the final drying phase, the product has reached a residual moisture
content of approx. 5%. Below this value, the product is generally not
damaged even if the power failure lasts for a longer period of time.
• If the product is in the main drying phase, we recommend aerating the
unit, removing the product, and storing it in a deep-freeze. The
defrosted condensate must be drained off prior to the next start.
Pos : 148 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 149 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020 Kein ausr eichendes Vakuum @ 25\ mod_1404984060039_68.doc x @ 184167 @ 3 @ 1
7.1.2
Insufficient vacuum
NOTE
The vacuum checks must be carried out when the ice condenser is frozen.
Pos : 150 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 151 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020-0010 Klei nflansc hverbi ndungen @ 25\ mod_1404984061086_68. doc x @ 184181 @ 4 @ 1
7.1.2.1
Small flange connections
Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
• Loosen the connection and place the centring ring (with sealing ring
inside) in a centred manner between the flange connections.
• Seal the connection with the clamping ring by tightening the wing nut.
• Ensure that the centring ring neither slips out of place nor gets jammed.