Chapter 3 Repair
77
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
1.
Remove the left and right casings
, as well as the
central casing
2.
Proceed to the dismantling of the wired motor
3.
Loosen the torquemeter-side coupling attachment screw (
A
).
4.
Remove the coupling from the torquemeter axle (
B
).
Note: Carefully collect the key (
C
) from the torquemeter axle.
For installing the coupling, take care to following points:
- The motor shaft shall be perfectly aligned with the torquemeter
axle and driving the whole line should be performed smoothly and
uniformly (manually turn the encoder wheel to check this).
- The coupling set screws shall be blocked using weak thread lock-
ing (Loctite 222).
- Calibrate the torquemeter
.
Operation 6-4 Dismantling/installing the motor coupling
Sheet: 1 / 1
Consumables:
Tools:
Cable ties
Loctite 222 (pink)
-
A
B
C
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...