ChoiceMMed MD300CF3 User Manual Download Page 1

Fingertip

 

Pulse Oximeter

 

USER MANUAL

 

Ver2.0CF3

 

 

General Description

 

Oxygen Saturation is a  percentage of Oxyhemoglobin (HbO

2

) capacity, compounded with oxygen, by  all  combinative hemoglobin 

(Hb)  capacity  in  blood.  In  other  words,  it  is  consistency  of  Oxyhemoglobin  in  blood.  It  is  a  very  important  parameter  for  the 
Respiratory  Circulation System.  Many  respiratory  diseases  can  result  in  oxygen  saturation being  lowered  in  human  blood. 
Additionally, the following factors can reduce oxygen saturation: Automatic regulation of organ dysfunction caused by Anesthesia, 
Intensive Postoperative Trauma, injuries  caused  by  some  medical  examinations. That  situation might  result  in light-headedness, 
asthenia, and vomiting. Therefore, it is very important to know the oxygen saturation of a patient so that doctors can find problems 
in a timely manner.

 

The  fingertip  pulse  oximeter  features  low  power  consumption, convenient  operation and  portability.  Place  one  fingertip  into  the 
photoelectric sensor for diagnosis and the pulse rate and oxygen saturation will appear on the display. It has been proven in clinical 
experiments that it also features high precision and repeatability.

 

Measurement Principle

 

Principle  of  the  oximeter  is  as  follows:  A  mathematical  formula  is  established  making  use  of  Lambert  Beer  Law  according  to 
Spectrum Absorption Characteristics of Reductive hemoglobin(RHb) and Oxyhemoglobin (HbO

2

) in glow and  near-infrared zones. 

Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity 
Pulse  Scanning  and  Recording  Technology, so  that  two  beams  of  different  wavelength of  lights  (660nm  glow  and  940nm  near 
infrared  light)  can  be  focused  onto  a  human  nail  tip  through  a  clamping  finger-type  sensor.  A  measured  signal  obtained  by  a 
photosensitive element, will be shown on the oximeter’s display through process in electronic circuits and microprocessor.

 

Diagram of Operation Principle

 

1. Red and Infrared-ray Emission Tube

 

2. Red and Infrared-ray Receipt Tube

 

 

Precautions For Use

 

1         

Before use, carefully read the manual.

 

2         

Operation of the fingertip pulse oximeter may be affected by the use of an electrosurgical unit (ESU).

 

 

The fingertip pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO

measurement. Verify 

that nothing is hindering the pulse measurement before relying on the SpO

measurement.

 

4         

Do not use the fingertip pulse oximeter in an MRI or CT environment.

 

 

Do  not  use  the  fingertip pulse oximeter in situations where alarms  are  required. The  device has  no  alarms.  It is  not  for 
continuous monitoring.

 

6         

Do not use the fingertip pulse oximeter in an explosive atmosphere.

 

 

The fingertip pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other 
methods of assessing clinical signs and symptoms.

 

8         

Check the pulse oximeter sensor application site every 4 hours to determine the positioning of the sensor and circulation

 

and skin sensitivity of the patient.

 

 

Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device in liquid. The device is not 
intended for sterilization.

 

10 

 

Follow local ordinances and recycling instructions regarding disposal or recycling of the  device and  device components, 
including batteries.

 

11       

This equipment complies with IEC 60601-1-2:2007 for electromagnetic compatibility for medical electrical equipment and/or

 

systems.  However,  because  of  the  proliferation of  radio-frequency transmitting equipment and  other  sources  of  electrical 
noise  in  healthcare  and  other  environments, it  is  possible  that  high  levels  of  such  interference  due  to  close  proximity  or 
strength of a source might disrupt the performance of this device.

 

12       

Portable and mobile RF communications equipment can affect medical electrical equipment.

 

Rx only: “Caution: Federal law restricts this device to sale by or on the order of a physician.”

 

 

Inaccurate measurements may be caused by

 

1         

Significant levels of dysfunctional hemoglobin (such as carbonyl - hemoglobin or methemoglobin);

 

2         

Intravascular dyes such as indocyanine green or methylene blue;

 

3         

High ambient light. Shield the sensor area if necessary;

 

4         

Excessive patient movement;

 

5         

High-frequency electrosurgical interference and defibrillators;

 

6         

Venous pulsations;

 

7         

Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line;

 

8         

The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia;

 

9         

The patient is in cardiac arrest or is in shock;

 

10       

Fingernail polish or false fingernails;

 

11       

Weak pulse quality (low perfusion);

 

12       

Low hemoglobin;

 

 

Product Properties

 

1       Operation of the product is simple and convenient.

 

2       The product is small in volume, light in weight and convenient to carry.

 

3       Power consumption of the product is low and the two AAA batteries can be operated continuously for 30 hours.

 

4       A low voltage warning will be indicated when battery voltage is low and normal operation of the oximeter might be influenced.

 

5       The product will automatically power off when there is no signal for longer than 8 seconds.

 

 

Intended Use

 

The fingertip pulse oximeter is a portable non-invasive device intended for spot checking of arterial hemoglobin oxygen saturation 
(SpO

2

) and pulse rate of adult and pediatric patient at home and hospital (including clinical use in internist/surgery, anesthesia and 

intensive care units). It is not for continuous monitoring.

 

 

Battery Installation

 

1.         Lift the battery door cover.

 

2.         Install  two  AAA  batteries  into  the  battery  compartment.

 

Match the plus (+) and minus (-) signs in the compartment. 

If  the  polarities are  not  matched, damage  may  be  caused 
to the oximeter.

 

3.         Close the battery door cover.

 

Notes:

 

       

Install the batteries with the correct polarity. Incorrect placement may cause damage to the bracket.

 

       

Please remove the batteries if the pulse oximeter will not be used for long periods of time.

 

 

Using the Lanyard

 

1.         Thread thinner end of the lanyard through the hanging hole.

 

2.        

Thread thicker end of the lanyard through the threaded end before pulling it tightly.

 

Warnings!

 

1. 

Keep the oximeter away from young children. Small items such as the battery door, battery, and 

lanyard are choking hazards.

 

2.      Do not hang the lanyard from the device’s electrical wire.

 

Product Accessories

 

1.         One lanyard

 

2.         Two AAA batteries

 

3.         One user manual

 

 

Operation Instructions

 

1          Install two AAA batteries according to the Battery Installation instructions.

 

2          Open the clamp as illustrated in the picture below.

 

3          Fully insert one fingertip into the silicone hole of the oximeter before releasing the clamp.

 

4          Press the switch button once on front panel.

 

5          Keep your finger still during measurement.

 

6          Read corresponding data from display screen.

 

7          Press the button again to toggle between six display modes.

 

After turning on the Oximeter, each time you press the power switch, the Oximeter will switch to another display mode. 
There are 6 display modes shown as follows:

 

 

1.  

        2.  

        3.  

 

 

4.  

        5.  

        6.  

 

 

Holding the power switch for longer than one second, will adjust the brightness of the oximeter. There are 10 levels of brightness. 
The default level is level four.

 

8          PR tone modulation function

 

9          Warning indication:

 

Indication range: SpO

2

: <90% PR Rate: <60 bpm    or >100bpm

 

Indication mode: Audible and Visible

 

Indication silence: W hen the audio indication occurs, the indication sound can pause for 30 seconds by pressing Power button 
and there is only the visual indication. By pressing the power button again, the audio indication will resume.

 

NOTE: 

This function is only valuable for audio indication, but not for pulse rate sound.

 

 

Front Panel

 

 

 

Patient pulse quality signals are indicated by bar graph. The bar is graded as 10 levels, if the strength is level 2 to 3, the pulse signal 
is inadequate.

 

Maintenance and Storage

 

1.         Replace the batteries in a timely manner when low voltage lamp is lighted.

 

2.         Clean surface of the fingertip oximeter before it is used in diagnosis for patients.

 

3.         Remove the batteries if the oximeter is not operated for a long time.

 

4.         It is best to store the product in -20

℃~

+55

 

and 

93% humidity.

 

5.         Keep in a dry place. Extreme moisture may affect oximeter lifetime and may cause damage.

 

6.         Dispose of battery properly; follow any applicable local battery disposal laws.

 

Cleaning the fingertip pulse oximeter

 

Please  use  medical  alcohol  to  clean  the  silicone  touching  the  finger  inside  of  oximeter  with  a  soft  cloth  dampened  with  70%

 

isopropyl alcohol. Also clean the being tested finger using alcohol before and after each test.

 

Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any openings in the device. Allow the oximeter to 
dry thoroughly before reuse.

 

A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor. Clinical testing is used to establish 
the  SpO

accuracy. The measured arterial hemoglobin saturation value (

SpO

2

) of the sensors is compared to arterial hemoglobin 

oxygen (

SaO

2

) value, determined from blood samples with a laboratory CO-oximeter. The accuracy of the sensors in comparison to

 

the   CO-oximeter   samples   measured   over   the  

SpO

2    

range   of   70  

–     

100%.   Accuracy   data   is   calculated   using   the

 

root-mean-squared (Arms value) for all subjects, per ISO 9919:2005, Medical Electrical Equipment

Particular requirements for the

 

basic safety and essential performance of pulse oximeter equipment for medical use.

 

The  use  life  of  the  device  is  five  years  when  it  is  used  for  15  measurements  every  day  and  10  minutes  per  one 
measurement. Stop using and contact local service center if one of the following cases occurs:

 

      

An error in the 

Possible Problems and solutions 

is displayed on screen.

 

      

The oximeter cannot be powered on in any case and not the reasons of battery.

 

  

There is a crack on the oximeter or damage on the display resulting readings cannot be identified; the spring is invalid; or the 

key is unresponsive or unavailable.

 

Specifications

 

1. Display Type

 

OLED display

 

2. SpO

2

 

Display range: 0-99% 
Measurement range: 70-99%

 

Accuracy: 70%-99%: ±3%;    0%~69% no definition

 

Resolution: 1%

 

3. Pulse Rate

 

Display range: 0~254BPM 
Measure range: 30-235 BPM

 

Accuracy: 30~99bpm, ±2bpm;    100~235bpm, ±2%

 

Resolution: 1BPM

 

4. Probe LED Specifications

 

 

 

Wavelength

 

Radiant Power

 

RED

 

660

±

2nm

 

1.8mW

 

IR

 

940

±

10nm

 

2.0mW

 

5. Power Requirements

 

Two AAA alkaline Batteries

 

Power consumption: Less than 40mA 

Low power indication:  

 

Battery Life: Two AAA 1.5V, 600mAh alkaline batteries could be continuously operated as long as 30 hours.

 

6.  Dimension 

L

ength:  60mm 

Width: 32mm

 