ChoiceMMed MD300C312 Operator'S Manual Download Page 1

 

OPERATOR’S 

MANUAL 

V1.0C3

 

Fingertip Pulse Oximeter

 

 

General Description   

Haemoglobin  Saturation  is  percentage  of  Oxyhemoglobin  (HbO2)  capacity,  compounded  with  oxygen,  by  all  combinativable 
haemoglobin (Hb) obin (HbO2) capacity in blood. In other words, it is consistence of Oxyhemoglobin in blood. It is a very important 
ecological parameter for Respiratory circulation System.  Many respiratory diseases can result in  haemoglobin Saturation being 
lowered in human blood. Moreover, the following factors can also lead to problems in oxygen supply, so that human haemoglobin 
saturation might be reduced: Automatic Organic Regulation Malfunction caused by Anesthesia, Intensive Postoperative Trauma, 
hurts resulted in by some medical examination and etc. In the situation, illnesses, such as light head, asthenia, vomitory and etc, 

might happen to patients and even endanger the patient’s life. Therefore, it is very important to know Hemoglobin saturation  of 
patient timely in clinical medical aspects. So that doctors can find problems in time. 

The fingertip pulse oximeter features in small volume, low power consumption, convenient operation and being portable. It is only 
necessary for patient to put one of his fingers into a fingertip photoelectric sensor for diagnosis, and a display screen will directly 
show measured value of hemoglobin Saturation. It has been proved in clinical experiments that it features in rather high precise 
and repeatability. 

Measurement principle 

Principle  of  the  oximeter  is  as  follows:  An  experience  formula  of  data  process  is  established  taking  use  of  Lambert  Beer  Law 
according  to  Spectrum  Absorption  Characteristics  of  Reductive  hemoglobin(R  Hb)  and  Oxyhemoglobin  (O2  Hb)  in  glow  and 
near-infrared  zones.  Operation  principle  of  the  instrument  is  Photoelectric  Oxyhemoglobin  Inspection  Technology  is  adopted  in 
accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm 
glow  and  940nm  near  infrared  light)  can  be  focused  onto  human  nail  tip  through  perspective  clamp  finger-type  sensor.  Then 
measured signal can be obtained by a photosensitive element, information acquired through which will be shown on two groups of 
LEDs through process in electronic circuits and microprocessor. 

Diagram of Operation Principle 

1.    Red and Infrared-ray Receipt Tube   

2.    Red and Infrared-ray Emission Tube 

Precautions for use 

Do not use the pulse oximeter in an MRI or CT environment 

Do not use the pulse oximeter in situations where alarms are required. The device has no alarms. 

Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere. 

The pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods 
of assessing clinical signs and symptoms. 

Check the pulse oximeter sensor application site frequently to determine the positioning of the sensor and circulation and 
skin sensitivity of the patient.   

Do not stretch the adhesive tape while applying the pulse  oximeter sensor. This may cause inaccurate  readings or skin 
blisters. 

Before use, carefully read the manual. 

The pulse oximeter has no SpO2 alarms; it is not for continuous monitoring, as indicated by the symbol.   

Prolonged use or the patient’s condition may require changing the sensor site periodically. Change sensor site and check 
skin integrity, circulatory status, and correct alignment at least every 4 hours. 

10 

Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing, or immersing the sensors in liquid may 
cause inaccurate readings. 

11 

Significant levels of dysfunctional hemoglobins (such as carbonxy- hemoglobin or methemoglobin) 

12 

Intravascular dyes such as indocyanine green or methylene blue 

13 

SpO2  measurements  may  be  adversely  affected  in  the  presence  of  high  ambient  light.  Shield  the  sensor  area  (with  a 
surgical towel, or direct sunlight, for example) if necessary. 

14 

Excessive patient movement 

15 

High-frequency electrosurgical interference     

16 

Venous pulsations 

17 

Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line 

18 

The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia 

19 

The patient is in cardiac arrest or is in shock 

20 

Fingernail polish or false fingernails may cause inaccurate SpO2 readings. 

Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components, 
including batteries. 

Product Properties 

Operation of the product is simple and convenient 

The product is small in volume, light in weight (total weight is about 50g including batteries) and convenient in carrying 

Power consumption of the product is low and the two originally-equipped two AAA batteries can be operated continuously 
for 30 hours.   

Low voltage warning will be indicated in visual window when battery voltage is so low that normal operation of the oximeter 
might be influenced. 

The product will automatically be powered off when no signal is in the product for longer than 8 seconds. 

Product Operation Scope 

The  fingertip  Oximeter  can  be  used  to  measure  human  Haemoglobin  Saturation  and  heart  rate  through  finger.  The  product  is 
suitable  for  use  in  family,  hospital  (including  clinical  use  in  internist/surgery,  Anaesthesia,  paediatrics,  intensive  care  and  etc.) 
Oxygen  Club,  social  medical  organizations,  physical  care  in  sports  (It  can  be  used  before  or  after  sports.  Operation  in  sport 
procedure is not recommended) and etc. 
The product is not suitable to monitor patient continuously. 

Operation Instructions 

Installing two AAA batteries into battery cassette before covering its cover. 

Nip the clamp as diagram 

Plug one of fingers into rubber hole of the Oximeter (it is best to plug the finger thoroughly) before releasing the clamp 

Press the switch button once on front panel. 

Your finger do not tremble during the Oximeter is working. Your body is not recommended in moving status 

Read correspondent datum from display screen. 

7

 

Six display modes 

 

After turn on the oximeter, each time you press the power switch, the oximeter will switch to another display mode, there 
are 6 display modes shown as follows: 

1

  2

  3

  4 

5

 

6

     

               

   

 

When  you  press  the  power  switch  for  a  long  time  (more  than  one  second),  the  brightness  of  the  oximeter  will  be  changed  by 
degrees, there are 10 levels on brightness; the default level is level four. 

Declaration: Please use the medical alcohol to clean the rubber touching the finger inside of Oximeter, and clean the test finger 
using alcohol before and after each test. (The rubber inside of the Oximeter belongs medical rubber, which has no  toxin, and no 
harmful to the skin of human being).   
When your finger is plugged into the Oximeter, your nail surface must be upward. 

8

 

PR tone modulation function

 

LCD display 

10 

Adjustable brightness 

11 

Alarm: 

Alarm range:     
SpO2:<90 
Pulse rate: 

<60 or >100 

Alarm mode: audio alarm and visual alarm 

Visual alarm:    The measuring value will be flashing when it is through over the alarm limitation.   
Audio alarm: The audible alarms can be heard if there is no silence. The buzzer will beep shortly for two times when the 
value excesses the alarm limitation. 

Alarm silence:   

When  the  audio  alarm  occurs,  press  the  button  and  then  the  Oximeter  will  enter  into  the  “alarm  silence  mode”  for  30 
seconds, and then there is only the visual alarm. This silence operating has only effect on the alarm sound not on the pulse 
rate sound.   

In the silence mode, press the button, and then the display mode will be changed.   

It will exit the silence mode automatically when a new alarm occurs or the value is in the normal range again. 

Brief Description of Front Panel

                                                                 

     

 

                                                                                                   

Low power indicator

 

SPO

2

 waveform                                                                                                                                                       

 

Bar graph SpO

2

 

Pulse rate

 

It is apparent the pulse rate bar graph display corresponds with pulse rate.   

Product Accessories 

1. One hangs lace 
2. Two batteries 
3. One user manual 

Battery Installation 

1. 

Put the two AAA batteries into battery cassette in correct polarities. 

2. 

Push the battery cover horizontally along the arrow shown as below: 

Notes: Battery polarities must be correctly installed. Otherwise, damage might be caused to device. 
Please put or remove batteries in right order, or is likely to damage the device bracket. 
Please remove the battery if the oximeter will not be used for long time. 

Hang Lace Installation 

1. 

Thread thinner end of the hang lace through the hanging hole. 

2. 

Thread thicker and of the lace through the threaded end before pulling it tightly. 

Maintenance and Storage 

1. 

Replace the batteries timely when low voltage lamp is lighted. 

2. 

Clean surface of the fingertip oximeter before it is used in diagnosis for patients. 

3. 

Remove the batteries inside the battery cassette if the oximeter will not be operated for a long time. 

4. 

It is best to preserve the product in a place where ambient temperatures 

–10-40

(14-104

) and humidity is 10%-80%. 

5. 

It is recommended that the product should be kept in a dry environment anytime. A wet ambient might affect its lifetime and 
even might damage the product. 

6. 

Please follow the law of the local government to deal with used battery.

 

Calibrating the pulse oximeter 

1. 

The functional tester cannot be used to assess the accuracy of the oximeter. 

2. 

Index 2 that made by Bioteck company is a function tester. Set Tech to 1, R curve to 2, then user can use this particular 
calibration curve to measure the oximeter. 

3. 

The  test  methods  used  to  establish  the  SpO

2

  accuracy  is  clinical  testing.  The  oximeter  used  to  measure  the  arterial 

haemoglobin oxygen saturation levels and these levels are to be compared to the levels determined from  arterial blood 
sampling with a CO-oximeter. 

Declaration:

   

EMC of this product comply with IEC60601-1-1-2 standard

 

The materials which user can come into contact is no toxicity and no action on tissues, comply with ISO10993-1,-5,-10. 

Detailed descriptions of product functions: 

Function 

Descriptions 

Function 

Descriptions 

Display Type 

LCD display

 

Declaration 

EMC  of  this  product  comply  with 
IEC60601-1-2 standard.

 

SpO2 

Measurement range:

70-99% 

Accuracy

±2%  on  the  stage  of  80%-99%; 

±3% on the stage of 70%-80%; 

Measurement 
Performance 

in 

Low 

Perfusion 

Condition 

required 

the 

test 

equipment 

(BIO-TEK  INDEX  Pulse  Oximeter 
tester) the pulse wave is available 
without failure when the simulation 
pulse wave amplitude is at 0.6%. 

Pulse Rate 

Measure range

30-235bpm 

Accuracy

30-99bpm 

 

±2bpm

 

100-235bpm      ±2%   
Pulse Intensity: Bargraph Indicator   

Interference 
Resistance 
Capacity  against 
Ambient Light 

Device work normally when mixed 
noise 

produced 

by 

BIO-TEK 

INDEX Pulse Oximeter tester 

Power 
Requirements

 

 

Two AAA alkaline Batteries 
Power consumption: Less than 30mA   
Low power indication: 
Battery  Life:  Two  AAA  1.5V, 
800mAh 

alkaline 

batteries 

could 

be 

continuously operated as long as 30 hours.

 

Environment 
Requirements 

Operation Temperature: 5-40

 

Storage Temperature: -10-40

 

Ambient Temperature: 

15%-80% in operation 

              10%-80% in storage

 

Dimension 

Length: 60mm          Width: 34mm        Height: 30mm        Weight: 50g (including two AAA batteries) 

PR tone modulation function 

Adjustable brightness 

Pulse signal strength indicator 

Guidance and manufacture’s declaration – electromagnetic emissions-for all EQUIPMENT and SYSTEMS

 

 

Guidance and manufacture’s declaration – electromagnetic emission 

The 

Pulse Oximeter 

is intended for use in the electromagnetic environment specified below. The customer of the user of the 

Pulse Oximeter 

should assure that it is used in such and environment. 

Emission test 

Compliance 

Electromagnetic environment 

– guidance 

RF emissions 

CISPR 11 

Group 1 

The 

Pulse Oximeter 

uses RF energy only for its internal function. Therefore, its RF 

emissions are very low and are not likely to cause any interference in nearby 
electronic equipment. 

RF emission 

CISPR 11 

Class B 

The 

Pulse Oximeter 

is suitable for use in all establishments, including domestic 

establishments and those directly connected to the public low-voltage power supply 
network that supplies buildings used for domestic purposes. 

Possible Problems and resolutions 

Problems 

Possible reason 

Solution 

SpO2 or PR 
can not be 
shown 
normally 

1. Finger is not plugged correctly 
2. 

Patient’s Oxyhemoglobin value is too low 

to be measured 

1..Retry by plugging the finger 
2. Try some more times, If you can make sure about no 

problem existing in the product. Please go to a hospital 
timely for exact diagnosis 

SpO2 or PR is 
shown 
unstably 

1. Finger might not be plugged deep enough 
2. 

Finger is trembling or patient’s body is in 

movement status 

1. Retry by plugging the finger 
2. Try not to move 

The Oximeter 
can not be 
powered on 
 

1. Power of batteries might be inadequate or 

not be there at all 

2. Batteries might be installed incorrectly 
3. The Oximeter might be damaged 

1. Please replace batteries 
2. Please reinstall the batteries 
3. Please contact with local customer service centre 

Indication 
lamps are 
suddenly off 

1. The product is automatically powered off   

when no signal is detected longer than 8 
seconds 

2. Power quantity of the batteries is started 

being inadequate 

1. Normal                 
2. Replace the batteries 

“Error3”    or   

“Error4” 
Displayed on 
screen 

1 .Low power   
2. Receiving tube being shielded or 

damaged together with broken connector. 

3. Mechanical Misplace for       

receive-emission tube   

4. Amp circuit malfunction. 

1. Change new battery   
2. Please contact with local customer service center 
3. Please contact with local customer service center 
4    Please contact with local customer service center 

“Error7” 
displayed on 
screen 

1. Low power   
2. Emission tube damaged.   
3. Current control circuit malfunction. 

1    Please change battery   
2    Please contact with local customer service center 
3    Please contact with local customer service center 

Symbol Definitions 

Symbol 

Definition 

Symbol    Definition   

 

The equipment type is BF 

 

Heart rate (BPM) 

 

Refer to  user manual before 

application 

   

 

Low power indication   

SpO

2

 

Hemoglobin saturation   

SN 

Serial No 

 

 