ChoiceMMed MD300-D Operator'S Manual Download Page 1

OPERATOR

MANUAL 

Ver 2.2D 

Fingertip Pulse Oximeter

 

 

General Description   

Haemoglobin  Saturation  is  percentage  of  Oxyhemoglobin  (HbO2)  capacity,  compounded  with  oxygen,  by  all  combinativable 
haemoglobin  (Hb)  obin  (HbO2)  capacity  in  blood.  In  other  words,  it  is  consistence  of  Oxyhemoglobin  in  blood.  It  is  a  very 
important  ecological  parameter  for  Respiratory  circulation  System.  Many  respiratory  diseases  can  result  in  haemoglobin 
Saturation being lowered in human blood. Moreover, the following factors can also lead to problems in oxygen supply, so that 
human  haemoglobin saturation might be  reduced: Automatic Organic Regulation  Malfunction caused by Anesthesia, Intensive 
Postoperative  Trauma,  hurts  resulted  in  by  some  medical  examination  and  etc.  In  the  situation,  illnesses,  such  as  light  head, 
asthenia, vomitory and etc, might happen to patients and even endanger the patient

s life. Therefore, it is very important to know 

Hemoglobin saturation of patient timely in clinical medical aspects. So that doctors can find problems in time. 
The fingertip pulse oximeter features in small volume, low power consumption, convenient operation and being portable. It is only 
necessary for patient to put one of his fingers into a fingertip photoelectric sensor for diagnosis, and a display screen will directly 
show measured value of hemoglobin Saturation. It has been proved in clinical experiments that it features in rather high precise 
and repeatability. 
 

Measurement principle 

Principle of the  oximeter is as follows: An experience formula of data process is established taking  use of Lambert Beer Law 
according  to  Spectrum  Absorption  Characteristics  of  Reductive  hemoglobin(R  Hb)  and  Oxyhemoglobin  (O2  Hb)  in  glow  and 
near-infrared zones. Operation principle of the instrument is Photoelectric Oxyhemoglobin Inspection Technology is adopted in 
accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm 
glow  and  940nm  near  infrared  light)  can  be  focused  onto  human  nail  tip  through  perspective  clamp  finger-type  sensor.  Then 
measured signal can be obtained by a photosensitive element, information acquired through which will be shown on two groups 
of LEDs through process in electronic circuits and microprocessor. 

Diagram of Operation Principle 

 
1.    Red and Infrared-ray Emission Tube 
2.    Red and Infrared-ray Receipt Tube 
 

Precautions for use 

1  Do not use the pulse oximeter in an MRI or CT environment 
2  Do not use the pulse oximeter in situations where alarms are required. The device has no alarms. 
3  Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere. 
4  The pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of 

assessing clinical signs and symptoms. 

5  Check the pulse oximeter sensor application site 

frequently

 to determine the positioning of the sensor and circulation and skin 

sensitivity of the patient.   

6  Do  not  stretch  the  adhesive  tape  while  applying  the  pulse  oximeter  sensor.  This  may  cause  inaccurate  readings  or  skin 

blisters. 

7  Before use, carefully read the manual. 
8  The pulse oximeter has no SpO2 alarms; it is not for continuous monitoring, as indicated by the symbol.   
9  Prolonged use or the patient

s condition may require changing the sensor site periodically. Change sensor site and check skin 

integrity, circulatory status, and correct alignment at least every 4 hours. 

10  Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing, or immersing the sensors in liquid may 

cause inaccurate readings. 

11  Significant levels of dysfunctional hemoglobins (such as carbonxy- hemoglobin or methemoglobin) 
12  Intravascular dyes such as indocyanine green or methylene blue 
13  SpO2 measurements may be adversely affected in the presence of high ambient light. Shield the sensor area (with a surgical 

towel, or direct sunlight, for example) if necessary. 

14  Excessive patient movement 
15  Venous pulsations 
16  Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line 
17  The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia 
18  The patient is in cardiac arrest or is in shock 
19  Fingernail polish or false fingernails may cause inaccurate SpO2 readings. 

Follow  local  ordinances  and  recycling  instructions  regarding  disposal  or  recycling  of  the  device  and  device 
components, including batteries. 
 

Product Properties 

1  Operation of the product is simple and convenient 
2  The product is small in volume, light in weight (total weight is about 50g including batteries) and convenient in carrying 
3  Power consumption of the product is low and the two originally-equipped two AAA batteries can be operated continuously for 

30 hours.   

4  Low voltage warning will be indicated in visual window when battery voltage is so low that normal operation of the oximeter 

might be influenced. 

5  The product will automatically be powered off when no signal is in the product for longer than 8 seconds. 

 

Product Operation Scope 

The fingertip Oximeter can be  used to  measure  human Haemoglobin Saturation and  heart rate through finger.  The product is 
suitable  for  use  in  family,  hospital  (including  clinical  use  in  internist/surgery,  Anaesthesia,  paediatrics, intensive  care  and  etc.) 
Oxygen  Club,  social  medical  organizations,  physical  care  in  sports  (It  can  be  used  before  or  after  sports.  Operation  in  sport 
procedure is not recommended) and etc. 
The product is not suitable to monitor patient continuously. 

 

Operation Instructions 

1  Installing two AAA batteries into battery cassette before covering its cover. 
2  Nip the clamp as diagram 
3  Plug one of fingers into rubber hole of the Oximeter (it is best to plug the finger thoroughly) before releasing the clamp 
4  Press the switch button once on front panel. 
5  Your finger do not tremble during the Oximeter is working. Your body is not recommended in moving status 
6  Read correspondent datum from display screen. 

7

  Six display modes 

 

 
After  turn  on  the  oximeter,  each  time  you  press  the  power  switch,  the  oximeter  will  switch  to  another  display  mode, 
there are 6 display modes shown as follows: 

1. 

    2. 

    3. 

    4. 

    5. 

    6. 

 

When you press the power switch for a long  time (more than one  second), the brightness of the oximeter will be changed  by 
degrees, there are 10 levels on brightness; the default level is level four. 
Declaration: Please use the medical alcohol to clean the rubber touching the finger inside of Oximeter, and clean the test finger 
using alcohol before and after each test. (The rubber inside of the Oximeter belongs medical rubber, which has no toxin, and no 
harmful to the skin of human being).   

 

When your finger is plugged into the Oximeter, your nail surface must be upward. 

 

Brief Description of Front Panel 

 
 
 

                                                                                       

Low power indicator

 

                                                                                     

 

SPO2 waveform 

It is apparent the heart rate bargraph display corresponds with pulse rate.   

 

Product Accessories 

1.  One hang lace 
2.  Two batteries 
3.  One user manual 

 

Battery Installation 

1.  Put the two AAA batteries into battery cassette in correct polarities. 
2.  Push the battery cover horizontally along the arrow shown as below: 
 
Notes: Battery polarities must be correctly installed. Otherwise, damage might be caused to device. 
Please put or remove batteries in right order, or is likely to damage the device bracket. 
Please remove the battery if the Oximeter will not be used for long time 

 

Hang Lace Installation 

1.  Thread thinner end of the hang lace through the hanging hole 
2.  thread thicker and of the lace through the threaded end before pulling it tightly 

 

Maintenance and Storage 

1.  Replace the batteries timely when low voltage lamp is lighted 
2.  Clean surface of the fingertip oximeter before it is used in diagnosis for patients 
3.  Remove the batteries inside the battery cassette if the Oximeter will not be operated for a long time 
4.  It is best to preserve the product in a place where ambient temperatures 

10-40

(14-104

) and humidity is 10%-80% 

5.  It is recommended that the product should be kept in a dry environment anytime. A wet ambient might affect its lifetime and 

even might damage the product. 

6.  Please follow the law of the local government to deal with used battery 
 

Calibrating the pulse oximeter 

1.  The functional tester cannot be used to assess the accuracy of the oximeter. 
2.  Index  2  that  made  by  Bioteck  company  is  a  function  tester.  Set  Tech  to  1,  R  curve  to  2,  then  user  can  use  this  particular 

calibration curve to measure the oximeter. 

3.  The  test  methods  used  to  establish  the  SpO2  accuracy  is  clinical  testing.  The  oximeter  used  to  measure  the  arterial 

haemoglobin  oxygen  saturation  levels  and  these  levels  are  to  be  compared  to  the  levels  determined  from  arterial  blood 
sampling with a CO-oximeter. 

 

Declaration:

   

EMC of this product comply with IEC60601-1-2 standard

 

The materials which user can come into contact is no toxicity and no action on tissues, comply with ISO10993-1,-5,-10. 

 

Detailed descriptions of product functions: 

1. 

Display Type: 

OLED display 

2.  SpO2: 

Measurement range:

70-99% 

Accuracy

±

2% on the stage of 80%-99%; 

±

2% on the stage of 70%-80%; 

3.  Pulse Rate: 

Measure range

30-235 BPM 

Accuracy

±

2 BPM or 

±

2%

larger

 

Pulse Intensity: Bargraph Indicator   

4. 

Power Requirements:

   

Two AAA alkaline Batteries 
Power consumption: Less than 40mA   

Low power indication:

 

 

Battery Life: 
Two AAA 1.5V, 600mAh alkaline batteries could be continuously operated as long as 30 hours.   

5.  Dimension: 

Length: 58mm 
Width: 32mm 
Height: 34mm 
Weight: 50g (including two AAA batteries) 

6.  Environment Requirements: 

Operation Temperature: 5-40

 

Storage Temperature: -10-40

 

Ambient Temperature:    15%-80% in operation 

10%-80% in storage   

7.  Declaration: 

EMC of this product comply with IEC60601-1-2 standard. 

 

8.  Measurement  Performance  in  Low  Perfusion  Condition:  required  the  test  equipment  (BIO-TEK  INDEX  Pulse 

Oximeter tester) the pulse wave is available without failure when the simulation pulse wave amplitude is at 6%.   

9.  Interference  Resistance  Capacity  against  Ambient  Light:  Device  work  normally  when  mixed  noise  produced  by 

BIO-TEK INDEX Pulse Oximeter tester   

 

Guidance and manufacture

s declaration 

 electromagnetic emissions-for all EQUIPMENT and SYSTEMS 

   

Guidance and manufacture

s declaration 

 electromagnetic emission 

The 

Pulse  Oximeter 

is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The  customer  of  the  user  of  the 

Pulse Oximeter 

should assure that it is used in such and environment. 

Emission test 

Compliance 

Electromagnetic environment 

 guidance 

RF emissions 
CISPR 11 

Group 1 

The 

Pulse Oximeter 

uses RF energy only for its internal 

function. Therefore, its RF emissions are very low and are not 
likely to cause any interference in nearby electronic 
equipment. 

RF emission 
CISPR 11 

Class B 

The 

Pulse Oximeter 

is suitable for use in all establishments, 

including domestic establishments and those directly 
connected to the public low-voltage power supply network 
that supplies buildings used for domestic purposes. 

 

Possible Problems and resolutions 

 

Problems 

Possible reason 

Solution 

SpO2 or PR can not 
be shown normally 

1. Finger is not plugged correctly 
2. Patient

s Oxyhemoglobin value is too 

low to be measured 

1..Retry by plugging the finger 
2. Try some more times, If you can make sure about no 

problem existing in the product. Please go to a hospital 
timely for exact diagnosis 

SpO2 or PR is shown 
unstably 

1. Finger might not be plugged deep 

enough 

2. Finger is trembling or patient

s body is in 

movement status 

1. Retry by plugging the finger 
2. Try not to move 

The Oximeter can not 
be powered on 
 

1. Power of batteries might be inadequate 

or not be there at all 

2. Batteries might be installed incorrectly 
3. The Oximeter might be damaged 

1. Please replace batteries 
2. Please reinstall the batteries 
3. Please contact with local customer service centre 

Indication lamps are 
suddenly off 

1. The product is automatically powered off   

when no signal is detected longer than 
8 seconds 

2. Power quantity of the batteries is started 

being inadequate 

1. Normal                2. Replace the batteries 

Error3

    or   

Error4

 Displayed on 

screen 

1 .Low power   
2. Receiving tube being shielded or 

damaged together with broken 
connector. 

3. Mechanical Misplace for       

receive-emission tube   

4. Amp circuit malfunction. 

1. Change new battery   
2. Please contact with local customer service center 
3. Please contact with local customer service center 
4    Please contact with local customer service center 

Error7

 displayed on 

screen 

1. Low power   
2. Emission tube damaged.   
3. Current control circuit malfunction. 

1    Please change battery   
2    Please contact with local customer service center 
3    Please contact with local customer service center 

 

Symbol Definitions 

 

Symbol    Definition 

Symbol 

 

 

The equipment type is BF

 

 

Heart rate (BPM) 

 

 

Refer to user manual before application

 

 

Low power indication 

SpO2 

Hemoglobin saturation

 

SN

 

Serial No 

 

 

PR Bargraph

 

Power 
Switch

 

SPO2

 

PR

 

 

 