ECO 6 Digital Color Doppler Ultrasound System
14
Operator Control Features:
The user should be aware that certain operator controls may affect the acoustic output. It is recommended to
use the default (or lowest) output power setting and compensate using Gain control to acquire an image. Other
than the output power setting in the soft-menu, which has the most direct impact on the power; the PRF, image
sector size, frame rate, depth, and focal position also slightly affect the output power. The default setting is
normally around 70% of the allowable power depending on the exam application mode.
Controls Affecting Acoustic Output
The potential for producing mechanical bioeffects (MI) or thermal bioeffects (TI) can be influenced by certain
controls.
Direct: The Acoustic Output control has the most significant effect on Acoustic Output.
Indirect: Indirect effects may occur when adjusting controls. Controls that can influence MI and TI are detailed
under the bioeffect portion of each control in the Optimizing the Image chapter.
Always observe the Acoustic Output display for possible effects.
Best practices while scanning
HINTS: Raise the Acoustic Output only after attempting image optimization with controls that have no effect
on Acoustic Output, such as Gain and STC.
WARNING:
Be sure to have read and understood control explanations for each mode used before
attempting to adjust the Acoustic Output control or any control that can affect Acoustic Output.
Use the minimum necessary acoustic output to get the best diagnostic image or measurement during an
examination. Begin the exam with the probe that provides an optimum focal depth and penetration.
Acoustic Output Default Levels
In order to assure that an exam does not start at a high output level, the system initiates scanning at a reduced
default output level. This reduced level is preset programmable and depends upon the exam icon and probe
selected. It takes effect when the system is powered on or New Patient is selected. To modify acoustic output,
adjust the Power Output level on the Soft Menu.
2.7
Battery Handling Instructions
Caution:
Read and observe the following warnings and precautions to ensure correct and safe use of
Li-ion batteries.
Do not immerse the battery in water or allow it to get wet.
Do not use or store the battery near sources of heat such as a fire or heater.
Do not use any chargers other than those recommended.
Do not reverse the positive (+) and negative (-) terminals.
Do not connect the battery directly to wall outlets or car cigarette-lighter sockets.
Do not put the battery into a fire or apply direct heat to it.
Do not short-circuit the battery by connecting wires or other metal objects to the positive (+) and negative
(-) terminals.