ECO 6 Digital Color Doppler Ultrasound System
127
Following all precautions for storage, use and disposal, prepare the germicide solution according to the
manufacturer’s instructions.
Place the cleaned and dried probe to contact with the germicide, being careful not to let the probe drop to
the bottom of the container and thus damage the probe.
After placing/immersing, rotate and shake the probe while it is below the surface of the germicide to
eliminate air pockets. Allow the germicide to remain in contact with the fully immersed probe. For high level
disinfection, follow the manufacturer’s recommended time.
Following all precautions for storage, use and disposal, prepare the germicide solution according to the
manufacturer’s instructions.
After removing from the germicide, rinse the probe according to the germicide manufacturer’s rinsing
instructions.
Flush all visible germicide residues from the probe and allow to air dry.
9.4
Service Responsibility
If users install, use and maintain the system fully according to CHISON’s installation manual, operation
manual and service manual, then CHISON ECO main unit has a life time of 5 years and CHISON ECO probes
have life time of 5 years after ex-work.
The warranty of the system and probes after ex-work is as the time in the warranty card.
The system is a precise electronic system. Only the CHISON’s authorized service engineer could replace the
defective parts. Any assembly, disassembly, handling, repair, or replacement by any other people may have
adverse impact on the safety and effectiveness of the systems and probes, and thus will reduce the life time of
the system and probes, and such systems and probes will not be covered by CHISON warranty after the above
improper handling. Standard maintenance must be performed by CHISON’s authorized service engineer
during the life time of the product.