![Chison ECO 1VET Operation Manual Download Page 118](http://html.mh-extra.com/html/chison/eco-1vet/eco-1vet_operation-manual_2602722118.webp)
ECO Series VET Digital Color Doppler Ultrasound System
103
Rooks VJ, Yancey MK, Elg SA, Brueske L. Comparison of probe sheaths for endovaginal sonography. Obstet.
Gynecol 1996;87:27-9.
Milki AA, Fisch JD. Vaginal ultrasound probe cover leakage: implications for patient care. Fertil Steril
1998;69:409-11.
Hignett M, Claman P. High rates of perforation are found in endovaginal ultrasound probe covers before and
after oocyte retrieval for in vitro fertilization-embryo transfer. J Assist Reprod Genet 1995;12:606-9.
Sterilization and Disinfection of Medical Devices: General Principles. Centers for Disease Control, Division of
Healthcare Quality Promotion. http://www.cdc.gov/ncidod/hip/sterile/sterilgp.htm (5-2003).
ODE Device Evaluation Information--FDA Cleared Sterilants and High Level Disinfectants with General
Claims for Processing Reusable Medical and Dental Devices, March 2003.
http://www.fda.gov/cdrh/ode/germlab.html (5-2003).
9.3
Probe Operation Instructions
For details on connecting, activating, deactivating, disconnecting, transporting and storing the probes.
9.3.1
Scanning the Patient
In order to assure optimal transmission of energy between the patient and probe, a conductive gel must be
applied liberally to the patient where scanning will be performed.
After the examination is complete, follow the cleaning and disinfecting, or sterilizing procedures as
appropriate.
9.3.2
Operating Transvaginal probe
The transvaginal probe is an endo-cavity probe, for the operation safety, please refer to “Care and
Maintenance” for cleaning and disinfection.
The temperature at the tip of the probe displays on the screen for monitoring. No temperature above 43º
C is
allowed. It also depends on the patient’s body temperature. When the temperature of probe tip exceeds 43º C,
the probe will stop working immediately to protect the patient.
Transvaginal probe should be used with FDA approved condom or probe cover. See the following instructions
to put the probe into the condom:
CAUTION
Some patients may be allergic to natural rubber or medical device with rubber contains. FDA suggests that
Summary of Contents for ECO 1VET
Page 2: ......