ECO Series VET Digital Color Doppler Ultrasound System
95
The probes are designed to be used only with this ultrasound system. Use of these probes on any other system
or a non-qualified probe may cause electrical shock or damage on the system/transducer.
9.2.2
Cleaning and Disinfecting
Place the probe into the solution of cleaning-disinfectant. Make sure not to immerse the probe into the liquid
beyond the immersion level given in the pictures below. Make sure that the probe is covered with the
cleaning-disinfectant up to the immersion level during the complete disinfection time.
For the recommended cleaning and disinfection time, please see your Operating Manual.
Rinse the probe with enough clean, potable water to remove all disinfectant residues.
Use a soft cloth to clean the cable and the user section of the probe with the cleaning disinfectant liquid.
Make sure that the surface of the probe and cable is wetted thoroughly with the cleaning-disinfectant.
Allow probe to air dry completely.
Reconnect the probe to the ultrasound console and place the probe into its holder.
Probe Immersion Levels
CAUTION
These transducers are not designed to withstand heat sterilization methods. Exposure to temperatures in excess
of 60 º
C will cause permanent damage. The transducers are not designed to be totally submerged in fluid, as
permanent damage will result if the entire transducer is submerged.
Probe Safety
Handling precautions
Ultrasound probes are highly sensitive medical instruments that can easily be damaged by improper handling.
Use care when handling and protect from damage when not in use. DO NOT use a damaged or defective probe.
Failure to follow these precautions can result in serious injury and equipment damage.