If no image or menu but image, please check following malfunction table. If the malfunction can’t be solved,
please contact with Sale Office, Customer Service department or Agents of manufacturer.
8.4
Malfunction Check
serial
number
malfunction
reason
Measures
1
Switch button lights
but power LED not
Battery
lose
efficacy,
Adapter works irregular
Check
the
connector
between cable and power
2
Power Led lights but
LED no image
The interval time is too
short to restart
Restart after 1 minute
3
LED display character
menu but no scan
image
Launch power, gain or STC
control errors
Not connect to probe or the
probe connection is not
correct
Device
is
in
freezing
condition
Control launch power , gain
or STC control
Make
sure
of
right
connection
Exit from freezing condition
4
Abnormal image
Exam mode errors
Image processing setting
errors
Whether Exam mode is
proper or not
Adjust image processing
setting or set it as default
5
Probe
works
improperly
1.The plug plugs loosely
2.Internal circuit protects
1.Extract the probe and
reinsert
2.Restart
6
No OB calculation
package menu
Do not select the OB
application
before
scanning.
Select the OB application
7
PRINT-button doesn’t
work
1.The connected printer is
approved
2.Printer power is not on
3.Printer is not connect well
1.Change
the
approved
printer
2.Turn on the printer
3.Connect the printer again