- 92 -
8-22. Fail Output Settings “FailOut”
Set the alarm operation performed at an activation of system related alarm (chart end, disconnection of input, SD card
error or low capacity, low backup battery level or other system error).
The SD card low-capacity alarm is activated when the free space on SD card decreases to 3% or lower.
The backup battery low-level alarm is activated when the voltage of backup battery for clock drops to 2.0V or lower.
The status information of other errors can be viewed by selecting “SysInfo” from the menu window.
Each alarm is turned off when the alarm condition is cleared or alarm operation is disabled in this setting (individual setting
available).
(1) Pressing
the
key displays the
menu window (list of setting items).
(2) Select
“FailOut”.
(3) Move the cursor to the parameter to be set
with the
▲
/
▼
/
/
keys.
(4) Press
the
key and check
the
check box of required items.
Also, set the desired relay No. for output
and circuit type.
(5) After completing the settings of this item,
move the cursor to Set .
(6) Press the
key to register the
settings (when chart recording is ON, a
setting change mark is printed). To cancel
the settings, press the
key.
Note: Actual windows are separated. Use the
▲
/
▼
keys to scroll and continue settings.