TROUBLESHOOTING CHART
PROBLEM
POSSIBLE CAUSE
SOLUTION
LCD display on
front panel does not
light up.
Plug incorrectly inserted
into socket.
Check that the socket is
working correctly.
Unit does not
function.
Power cable incorrectly
inserted
into
the
connector on the rear of
the unit.
Insert the plug correctly
into the socket.
Cable is worn, damaged
or blocked.
Replace the worn out or
damaged power cable.
The switch on the rear of
the unit is turned off.
Turn on the switch.
Fuses missing, blown or
blocked.
Replace
any
missing,
blown
or
interrupted
fuses.
Electronic control circuit
does not work
Contact
CHINESPORT
SpA Service centre .
No power reaching the
socket.
LCD display on
front panel does not
light up.
Presence
of
faulty
components on electronic
control card.
Contact
CHINESPORT
SpA Service centre .
Some of the buttons
on the front control
panel do not
function correctly
Faulty keys or buttons.
Contact
CHINESPORT
SpA Service centre .
Electronic control circuit
malfunction.
The unit lights up
but does not emit
Parameters
not
set
correctly.
Check that the parameters
have been set correctly.
energy
Laser source does not
function or has run out.
Check
laser
source
emission is operating.
Faulty components on
electronic control circuit.
Contact
CHINESPORT
SpA Service centre .
Faulty supply on laser
circuit.
The unit functions
normally but with a
significant decrease
in efficiency of
treatment.
Faulty or depleted laser
source.
Contact
CHINESPORT
SpA Service centre .
Possible break down in
power generator circuit
of the unit.
The equipment
starts up, or seems
to work normally,
but there is no
emission.
No safety key or the
interlock circuit is open.
Insert the DIN safety key
into the back socket or
reset the safety conditions.