![China Electronics Technology Instruments AV3672 Series User Manual Download Page 243](http://html1.mh-extra.com/html/china-electronics-technology-instruments/av3672-series/av3672-series_user-manual_2601790243.webp)
7 Calibration
7.10 Electronic Calibration
239
button.
d)
[Port Matching]
drop-down menu
The port selection part includes the vector network port and fixture port. The [Fixture port] drop-down menu is not
displayed in the default mode. The selected mode and network embedding/de-embedding type will be displayed in
the schematic diagram. Select the corresponding port relationship to determine the number of the embedding/de-
embedding port and the corresponding relationship to the port for file import.
e)
[Enable on/OFF]
check box
Select this check box to enable the fixture embedding/de-embedding function.
f)
[OK]
button
Click it to make the settings in the dialog box effective for the current active channel.
2) Steps of 4-port embedding/de-embedding
Menu path:
[Cal] > [Fixtures]
. Click
[4-port Embed/De-embed...]
in the
[Fixtures]
sub-menu. The
[4-port
Embed/De-embed]
dialog box will appear.
Tick the
[Enable on/OFF]
button in the dialog box to enable the four-port fixture embedding/de-embedding
function.
Click the
[Mode selection]
drop-down menu and select the embedding/de-embedding network model. The vector
network analyzer is configured with three network models (
Mode A, B and C
).
Select the embedding/de-embedding type of Network 1 in the drop-down menu of
Network 1
(
Network 2
is
enabled in
Mode C
).
If
embedding
or
de-embedding
is selected, the
[Load File]
button will be available. Click the
[Load File]
button
and select the
S4P
file to be used in the pop-up dialog box. The file name will be displayed in the lateral square box.
Click
[OK]
to complete the setting.
7.10 Electronic Calibration
Electronic calibration is a process to obtain system errors by means of standard measurements of the electronic
calibration kit through the network analyzer and calculation according to the fixed value of the electronic
calibration kit. The accuracy of electronic calibration is close to that of mechanical calibration. At the same time,
artificial errors arising from connection of the standard calibration kit in calibration of the vector network analyzer
can be reduced, thus improving the calibration accuracy, calibration velocity and calibration result consistency and
reducing the losses of the test port and expensive test cable.
Each electronic calibration kit is configured with a unique S-parameter saved in the internal memory. In the
calibration process, the vector network analyzer can recall the standard S-parameter of the electronic calibration kit
to obtain the vector error coefficient by means of complex measurement and calculation.
Compared with mechanical calibration, electronic calibration has obvious advantages:
1) High velocity and automatic operation. All the calibration can be completed by one connection, thus reducing
the times of calibration kit connection, menu button selection operations, operation time, connector wear and costs
for maintenance of the test port connector of the test instrument and the calibration kit connector.
2) Low risk and high measurement accuracy. The requirements for operators are low, and the possibility of errors
caused by operators is reduced, which is conducive to the promotion of the network analyzer in the military,
research and production applications.
3) New technology and high quality. New calibration and error correction technologies are applied to make the
accuracy higher than that of mechanical calibration.
Summary of Contents for AV3672 Series
Page 3: ......
Page 4: ...AV3672 Series Vector Network Analyzer Contents...
Page 5: ......
Page 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Page 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Page 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Page 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Page 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Page 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Page 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Page 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Page 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Page 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Page 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Page 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Page 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Page 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Page 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Page 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Page 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Page 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Page 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Page 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Page 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...