![China Electronics Technology Instruments AV3672 Series User Manual Download Page 202](http://html1.mh-extra.com/html/china-electronics-technology-instruments/av3672-series/av3672-series_user-manual_2601790202.webp)
6 Measurement Optimization
6.7 Increase of Data Point Number
198
Fig. 6.17 Setting of Intermediate Frequency Bandwidth
6.7 Increase of Data Point Number
The data point or “point” refers to the sampling results of next value after one stimulus. You can define the number
of measurement points within one sweep.
➢
The trace can be swept through a large number of data points so as to better define the device response.
➢
The majority of data processing is done between adjacent points. One “sweep” refers to the measurement
results of a series of continuous data points, corresponding to the stimulus value within one sequence.
➢
The default number of points of each sweep is 201.
The sweep time of the analyzer can be changed according to the number of sweep points and the scale. However,
the total measurement period will not change according to the number of sweep points and the scale. Instead, it is
affected by other factors, such as the flyback time, data calculation, formation time, etc. Refer to the technical
specifications for the influence of the point number and other settings on the sweep time.
Prompt
Use of data points
➢
In order to obtain the best trace resolution, use the maximum number of data points.
➢
In order to realize rapid throughput, use the minimum number of data points, which can help to achieve the
acceptable accuracy.
➢
In order to find the optimal number of data points, find a value with measurement results subject to no
significant difference from the required number of measurement points.
➢
In order to ensure the accurate measurement calibration, the number of calibration points should be the same
as that of measurement points.
1) Specific operations
Menu path:
[Stimulus] > [Sweep] > [Number of Points]
.
Directly click the required point number in the
[Number of Points]
sub-menu, or click
[Custom...]
, enter the value
in the
[Number of Points]
dialog box, or select the value with the arrow.
Summary of Contents for AV3672 Series
Page 3: ......
Page 4: ...AV3672 Series Vector Network Analyzer Contents...
Page 5: ......
Page 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Page 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Page 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Page 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Page 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Page 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Page 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Page 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Page 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Page 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Page 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Page 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Page 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Page 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Page 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Page 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Page 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Page 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Page 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Page 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Page 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...