4 Measurement Setting
4.6 Trigger Mode
100
Fig. 4.18 Trigger Setting Dialog Box
4.6.3 [Trigger] Dialog Box
1) Trigger setting options
a) Trigger source zone
The source of the channel trigger signal is determined by trigger source setting of the trigger source zone. The
effective trigger signal is generated when sweep is not done by the vector network analyzer. Three trigger sources
can be selected:
internal
,
manual
and
external
. The set trigger source will be used as the trigger source of all
channels.
➢
[Internal] (default)
1) Trigger signal generation is controlled automatically by the analyzer.
2) One trigger signal is generated immediately after one measurement.
➢
[Manual]
The manual trigger signal is generated in the following methods.
1) Click the
[Manual]
button in the
[Trigger]
dialog box.
2) Click the
[Manual
item in the
[Trigger]
sub-menu under the
[Stimulus]
menu of the menu bar.
➢
[External]
1) The trigger signal is input by
[External]
of the BNC connector on the rear panel.
2) The signal is TTL level.
3) Level trigger: the high or low level can be set in the
[External]
dialog box.
4) The pulse width is 1μs at and must not be longer than the sweep time (otherwise, the trigger may be enabled
several times).
b) Range zone
Summary of Contents for AV3672 Series
Page 3: ......
Page 4: ...AV3672 Series Vector Network Analyzer Contents...
Page 5: ......
Page 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Page 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Page 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Page 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Page 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Page 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Page 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Page 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Page 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Page 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Page 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Page 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Page 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Page 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Page 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Page 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Page 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Page 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Page 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Page 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Page 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...