Check-line AD-100 Operating Instructions Manual Download Page 1

Operating Instructions

CHECK•LINE

®

                       BY ELECTROMATIC 

Analog Durometer 

Model AD-100

Summary of Contents for AD-100

Page 1: ...Operating Instructions CHECK LINE BY ELECTROMATIC Analog Durometer Model AD 100 ...

Page 2: ... 1 0 Introduction 2 1 1 Unpacking 1 2 Complete Kit 2 0 Testing Procedure 3 3 0 Memory Pointer 4 4 0 Specifications 5 4 1 Conforms to International Standards 4 2 Dimensional Drawings 5 0 Calibration Repair 7 6 0 Warranty 8 ...

Page 3: ... are available A B C D DO O OO The AD 100 conforms to ASTM D 2240 DIN 53505 ISO 868 ISO 7619 1 1 Unpacking Unpack the durometer and inspect it for any shipping damage Notices of defect must be filed immediately in writing at the latest within 10 days on receipt of the goods 1 2 Complete Kit The AD 100 is supplied as a complete kit including the durometer instruction manual and NIST traceable calib...

Page 4: ...of the gauge firmly against the specimen but not so firmly as to imbed the foot into the surface of the material 3 Maintain pressure for 2 to 3 seconds The dial hand gives the reading in durometer points Important After the initial reading has been noted continue maintaining pressure for several more seconds Creep or cold flow of the specimen if present will be evident by receding action of the di...

Page 5: ...imen is released and the measurement pointer returns to zero the Memory Pointer remains in position marking the last measurement To return the Memory Pointer to the zero position prior to taking another measurement rotate the knurled knob on the face of the instrument counter clockwise until the pointer has been reset Memory Point er Knurled Knob ...

Page 6: ...se ranges be omitted Measuring Range 0 to 100 units Accuracy 1 durometer unit A B C D O DO 2 durometer units OO Dial Size 1 97 50mm diameter Dimensions 5 04 x 2 24 x 1 18 128 x 57 x 30 mm Weight 7 ounces 200 grams Warranty 1 year 4 1 Conformance to International Standards Model ASTM D2240 DIN 53505 ISO 868 ISO 7619 AD 100 A AD 100 B AD 100 C AD 100 D AD 100 DO AD 100 O AD 100 OO ...

Page 7: ... 6 0 71 in 18mm 0 98 in 25mm 0 79 in 20mm 2 24 in 57mm 5 04 in 128mm 8oz 1 00 in 25 4mm 4 1 Dimensional Drawings ...

Page 8: ... 7 5 0 Calibration and Repair All Check Line durometers requiring calibration and or certification should be sent directly to ELECTROMATIC Equipment Co Inc ...

Page 9: ...E RESULTING FROM THE USE OF SUPPLIED EQUIPMENT WHETHER SUCH DAMAGE OCCURS OR IS DISCOVERED BEFORE UPON OR AFTER REPLACEMENT OR REPAIR AND WHETHER OR NOT SUCH DAMAGE IS CAUSED BY MANUFACTURER S OR SUPPLIER S NEGLIGENCE WITHIN ONE YEAR FROM INVOICE DATE Some State jurisdictions or States do not allow the exclusion or limitation of incidental or consequential damages so the above limitation may not a...

Page 10: ...Distributed by ABQ Industrial LP USA Tel 1 281 516 9292 888 275 5772 eFax 1 866 234 0451 Web https www abqindustrial net E mail info abqindustrial net ...